![Samwontech TEMI2000 Series Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/samwontech/temi2000-series/temi2000-series_communications-manual_3790127046.webp)
SAMWONTECH
1st Edition of TEMI2000_Series IM : Dec. 15. 2008 Page 44 / 58
7.16.2 How to READ program PATTERN
▶ Below describes process step to read programmed PATTERN profile in TEMI2500.
① Write PATTERN number to read profile in D1701
② Write SEGMENT number to read profile in D1702
③ Write READ TRIGGER('2') in D1707
④ Stay until reading D1708 response DONE('1')
⑤ Reads profile(TSP of temp.& humi. process time etc.)in D1725~D1733
Above process step ① ~ ⑤ is used to read ‘ONE SEGMENT’profile among all in programmed pattern.
To read many segments, reiterate ① ~ ⑤ process step by changing segment number.
Setting ‘0’in D1702 at process step ② will read profile in D1736~D1757.