SAMWONTECH
4th Edition of TEMI880 IM : Jan. 29. 2007 Page 70 / 119
○
11
HUMI SOAK
SIGNAL
Set relay number for the humidity
Soak signal output.
▶ When Fix running, status lamp of the
running screen is display but relay
output is not occurred.
▶ When Program running, status lamp
of the running screen and relay is
“ON” during Soak time minus setting
time.
○
13
DRAIN
SIGNAL
Set relay number for the drain water
of the chamber.
▶ POWER ON
→ PROG STOP (Relay “ON”)
▶ PROG RUN
→ Temp. SP != DRY LIMIT Range
(Relay “ON”)
→ Humi. SP is 0.0% (Relay “ON”)
▶ RUN -> STOP → “HUMI RUN
SIGNAL” STOP (Relay “ON”)
○
14
DRAIN
RANGE LOW
Display DRAIN range low of the
ON/OFF operation.
▶ A case DRY
LIMIT LOW of
2.12.1.3 Input Display Screen
is changed, it is changed.
☞ Not edit.
○
15
DRAIN
RANGE HIGH
Display DRAIN range high of the
ON/OFF operation.
▶ A case DRY
LIMIT HIGH of
2.12.1.3 Input Display Screen
is changed, it is changed.
☞ Not edit.
○
16
ERROR
SIGNAL
Set relay number for the error signal
output.
▶ when DI error(DI2 ~ DI8) is occurred,
Relay status is “ON” while setting
time.
○
17
PATTERN
END SIGNAL
Set relay number for the pattern end
signal output.
▶ when program pattern is ended,
Relay status is “ON” while setting
time.
○
18
1st REF.
SIGNAL
Set relay number for the first REF.
signal output.
▶ A case where Inner Signal1(IS1)
output is occurred, Relay status is
“ON” after setting time(second).
○
19
2nd REF.
SIGNAL
Set relay number for the second
REF.
signal output.
▶ A case where Inner Signal1(IS1)
output is occurred, Relay status is
“ON” after setting time(minute).
☞ A case where setting relay number is duplicated, an occasion what kind of signal output is occurred, relay
is operated.(“OR” Condition)