2.2.6 Choose if a additional step is required
New
_
Step
_
?
______
Yes No
+/-
Select desired answer
accept
If
Yes
selected: step number will increment with 1
Ä
2.2.7
If
No
selected:
Ä
2.2.12
2.2.7 Set Temperature – Step 2
Set
_
Temperature
_
04/02
____
110.0°C
+/-
change desired value
confirm the value
Ä
2.2.8
ESC
restores the old value or
if in step 2 and higher
returns to
Ä
2.2.6
but the actual step will not be saved !
Note
: Display 04/02 means actual program/step number
2.2.8 Set Gradient – Step 2
Gradient
04/02 2.0°C/Min
+/-
change desired value
confirm the value
Ä
2.2.9
ESC
restores the old value or
if in step 2 and higher
returns to
Ä
2.2.6
but the actual step will not be saved !
2.2.9 Set Holding Time – Step 2
Hold
_
Time
___
04/02
_
H:M
_
10:30
+/-
Change desired value
confirm value and skips to the next input field (HH
Ä
MM) or
stores the time and go to
Ä
2.2.10
ESC
restores the old value and skips back one input field (M
Ä
H)
or
if in step 2 and higher
returns to
Ä
2.2.6
but the actual step
will not be saved !
Note
: A value of 0:00 means endless holding time
2.2.10 Set Fan Speed - Step 2
Fan
_
Speed
04/02
______
100%
+/- change desired value
confirm the value
Ä
2.2.11
ESC
restores the old value or
if in step 2 and higher
returns to
Ä
2.2.6
but the actual step will not be saved !
Copyright© SalvisLab / Edition 191107