Salutron Mini GlossTector 60 Скачать руководство пользователя страница 6

5 Overview  

 

 

1 Display 

2 Main button  

3 String position 

4 USB input  

5 Holder 

6 Calibration board 

 

6 Getting started 

6.1 Turn on  

Turn on the mini 

Gloss

Tector

 60° by shortly pressing the 

main button

.  

 

6.2 Calibration 

When the instrument is placed in the 

holder

 it automatically enters calibration mode. 

The display shows 

CAL

. When the 

main button

 

gets pressed the instrument calibrates to 97.0 

Gu.  Its  advisable  to  calibrate  the  instrument 

every  time  it  gets  turned  on  or  after  a  long 

series  of  measurements  to  ensure  optimal 

precision.  Note  that  on  the  gauge  and  on  the  holder  one  can  see  an  arrow.  When 

placing the gauge in the holder make sure both arrows stand in line to each other. 

 

 

1 Calibration 

2 Test 

3 Statistic 

4 Geometry 

5 Battery 

6 Standard Deviation  

7 Average  

8 Error  

9 Result   

Содержание Mini GlossTector 60

Страница 1: ...Mini GlossTector 60 Operations Manual ...

Страница 2: ......

Страница 3: ...4 2 SAFETY 4 3 GENERAL INFORMATION 4 4 POWER SUPPLY 5 5 OVERVIEW 6 6 GETTING STARTED 6 6 1 TURN ON 6 6 2 CALIBRATION 6 6 3 TESTING 7 6 4 TURN OFF 7 7 STATISTICS 7 9 DATA TRANSFER 8 10 ERROR NOTIFICATIONS 8 11 TECHNICAL DATA 8 12 STANDARDS 9 ...

Страница 4: ...d of time 3 General information Before using the mini GlossTector 60 we strongly advice you to read the instructions manual After receiving a delivery we ask you to check whether everything is included and not damaged The instrument consists of sensitive optical and electronic precision part Prevent it from being dropped Do not let any object to get into the aperture Do not expose the instrument t...

Страница 5: ...ply The instrument gets supplied with one 1 5V AAA LR03 alkaline battery Normally one battery should last for about 10000 measurements When the battery voltage approaches the required minimum level a notification will appear on the display Changing Battery Open the battery cover on the back of the instrument Insert a new battery into the battery box paying attention to the positive and negative po...

Страница 6: ...n mode The display shows CAL When the main button gets pressed the instrument calibrates to 97 0 Gu Its advisable to calibrate the instrument every time it gets turned on or after a long series of measurements to ensure optimal precision Note that on the gauge and on the holder one can see an arrow When placing the gauge in the holder make sure both arrows stand in line to each other 1 Calibration...

Страница 7: ... main button is pressed for 2 seconds 7 Statistics To enter the statistics mode press and hold the main button for about 2 seconds while its not placed in the holder The instrument will display N0 In statistics mode is possible to take up to 10 samples and the instrument will show the average and standard deviation of those data points Simply place the instrument at the points one wants to measure...

Страница 8: ...eu downloads 10 Error notifications Calibration error Clean the calibration standard and the lens carefully using ethanol or cleaning alcohol If the calibration should still not work contact Salutron Messtechnik GmbH our technical assistance is happy to help Wrong results If the gauge should display wrong results calibrate it and try again If the results should still be wrong contact Salutron Mess...

Страница 9: ...20 degrees 60 degrees and 85 degrees ASTM D 523 Standard test method for secular gloss DIN 67530 The reflect meter as an aid in evaluating gloss on level paint and plastic surfaces ISO7668 Anodized aluminum and aluminum alloys measurement of secular reflectance and secular gloss at angles of 20 60 and 85 ...