31
Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
value of previous measured trace data.
[Min. Hold]
Menu description: to set the current trace processing method as minimum hold, i.e., the new trace data will be assigned with minimum
value of previous measured trace data.
[Blank]
Menu description: Clear all displayed traces. Trace save in register will not be deleted.
[View]
Menu description: Hold current trace and not refresh. Easy for checking.
[Detector]
Pop up soft menus related to setting the detector type for the currently selected trace, including [Auto], [Normal], [Pos Peak], [Sample],
[Neg Peak].
[Auto]
Menu description: to set the detector command for the currently selected trace as Auto. That means, the signal/spectrum analyzer
will automatically select most suitable detector based on the parameter setting of currently selected trace, including trace
processing method, trace average type, noise marker function, etc.
[Normal]
Menu description: to set the detector command for the currently selected trace as Normal. i.e., display the measurement results of
positive peak and negative peak at the same time to achieve display effect similar to the analog instrument.
[Positive Peak]
Menu description: to set the detector command for the currently selected trace as Positive Peak. i.e., take the maximum value
from sampling data assigned to each trace point.
[Sample]
Menu description: to set the detector command for the currently selected trace as Sample. i.e., take the last value from sampling
data assigned to each trace point.
[Negative Peak]
Menu description: to set the detector command for the currently selected trace as Negative Peak. i.e., take the minimum value
from sampling data assigned to each trace point.
[Trace Operation]
Menu description: to start up trace mathematical operation and set the methods and rules. The soft menus include [1<-->2], [2-DL->2],
[2<-->3], [1->3], [2->3].
[1<-->2]
Menu description: Exchange the data in trace register 1 and trace register 2 and display
[2-DL->2]
Menu description: Data in trace register 2 minus current trace and save the result in trace register 2 and display.