23
Fig.17 Interface of the automatic test of indexes
(3)
Fill in the communication parameters in the communication parameter setting column, including the IP
address of the spectrometer, signal source address (S1465H), and power meter address. The address of the signal
source and power meter can be inquired through System menu. If the GPIB number of the signal source is 19, no
modification is required; similarly, if the GPIB number of the power meter is 13, no modification is required. The
IP address of the spectrometer can be inquired through
【
System
】
- [Network Configuration].
Attention: The IP address of the S3302 spectrometer and PC shall be located on the same network segment.
If they are different, the IP of the computer or spectrometer shall be set to make it within the same network
segment (for example, if the IP of the spectrometer is 172.141.11.202, the IP of the computer shall be set to
172.141.11.xx).
Click the indexes to be tested, and click Start Test. Attention:
①
In case of measurement and display of the average noise level and remaining response, a 50Ω load shall be
connected at the RF port. The indicators including test frequency range, third-order intercept point, 1 dB gain
compression and standing-wave ratio can only be manually tested.
②
In case of initial measurement of the reference level, scale fidelity and total level uncertainty, the power
meter shall be connected for calibration. No recalibration is needed when another spectrometer is tested.
(4)
After the test of all indexes is completed, click Save Data to save the test results in the form of word files.
- End of Document -