4.6 Other common faults and treatment methods
Fault Description
Causes
Solution
Cannot be started normally.
The battery has run down.
Charge the battery and observe the switch key
lamp. If the red light is displayed, continue
charging. Otherwise, contact the supplier.
Cannot be charged properly.
The use environment does not meet the
charging conditions.
Charging environment is 0
℃~
50
℃
.
Battery contact is poor.
Open the battery bin cover and check whether
the battery contacts are off.
Battery or internal circuit problems.
Replace the battery.
Normal curve cannot be
measured
Parameters setting are incorrect.
Reset the correct test parameters.
The output connector is polluted.
Clean optical output connector.
The output connector is damaged.
Replacement of output connector.
Optical output connectors do not match.
Replace the matched connector, the locating pin
of jumper joint should be aligned with the
flange notch and tightened.
The test curve has big burrs,
waveform unevenness.
The output connector is incorrect.
Reconnect the appropriate output interface.
Pulse width setting is too small.
Increase the test pulse width.
Saturation (flat top)
Pulse width setting is too large.
Reduce the test pulse width.