Document Reference: ENG-UMAN-000731
Revision: 5.00
Page 47 of 61
Settings
Go to the Settings Menu (see Section 9.8)
Programming
Go to the CLS programming screen.
9.8
Settings Menu
Touch the “Settings icon” for additional CPA functions.
Navigation Menu
Open the Navigation Menu.
Stop
Stop stimulation
Engineering
Information
Used by Saluda Medical staff to obtain
device related information.
Set Defaults
Set Settings page values to defaults
Pulse Shape
Biphasic - Positive first, Negative first.
Triphasic - Positive first, Negative first.
Interphase Gap
20 – 500 μs
Min Inter-Stimulus
Interval
Interval of stimulation set pulses
Auto (minimum permitted ISI),
1400 – 50000 μs
Max Stimulation
Voltage
7.5 – 15 V
Max Stimulation Ramp
Rate
Only applies when connected to CPA.
0.05 – 5.0 mA/s
ERT Amplifier Setting
ERT amplifier gain setting.
High gain (± 1.8 mV), Low gain (± 7.2 mV)
Sampling Delay
100 – 1000 μs
Max FBV Target Ramp
Rate
Only applies when connected to CPA.
1 – 200 μV/s
Factory Reset
Restore devices to default factory settings.
For use by Saluda Medical Field Clinical
Engineer (FCE).
Reset of Last Resort
For use by FCE.
Enter Stock Mode
Stop all CLS functions to conserve battery.
Use the Charger to leave stock mode.
Visit Name
Set label name for programming visit