Document Reference: ENG-UMAN-000730
Revision: 6.00
Page 26 of 40
4.
Line up the end of lead adapter plug with the end of the socket in the lead adapter
cable at an angle to ensure the connector pins do not get bent. Push the lead
adapter plug into the socket on the lead adapter cable until it clips into place.
•
The lead adapter cable is labelled “1” on one side for electrodes 1-12 and “2”
on the other side for electrodes 13-24.
Figure 10.5: Placing the lead adapter plug into the socket on the lead adapter cable.
5.
To connect a second lead to the lead adapter, repeat the above steps 1 to 4.
Figure 10.6: Placing a second lead adapter into the lead adapter cable.
Connect the lead adapter cable to the lead adapter extension
1.
Plug the lead adapter extension into the proximal connector of the lead adapter
cable until you feel a gentle click.
2.
Pass the proximal end of the lead adapter extension out of the sterile field to the
programming clinician.
Connect the lead adapter extension to the eCLS
This section is to be completed by the programming clinician outside of the sterile field.