background image

 

 

 

.Key Definition summary: 

KEY MODE 

DEFINITION 

Normal weighing mode 

Enter or exit HOLD mode; output the 
data as per P4,P5,P6 setting 

Setup mode/Calibration 

mode 

Shift the flashed position from right to 
left 

HOLD 

/PRINT 

Displaying A/D code or input 

voltage mode 

No Function 

Normal weighing mode 

Choose weight units, refer P8, P9, P10, 
P11 setting and Table3,Table4.  

Setup mode/Calibration 

mode 

Change the digit on flashed position and 
click this button to add 1.  

UNIT 

Displaying A/D code or input 

voltage mode 

Choose the weight inner code or input 
working voltage to be displayed.  

Normal weighing mode  Tare the weight 

Setup mode/Calibration 

mode 

Confirm the displayed parameters or 
input data, and go to next step  

TARE 

Displaying A/D code or input 

voltage mode 

Choose filtered or un-filtered weight 
A/D data 

ON/OFF/

ZERO 

Normal weighing mode  Zero function or Power off the scale.  

26 

 

Setup mode/Calibration 

mode/Displaying A/D code 

or input voltage mode 

Exit and back to normal work mode 

ON/OFF/

ZERO 

HOLD 

/PRINT

Normal weighing mode     

(more than 3s) 

Go to Show A/D code or input working 
voltage of indicator mode 

ON/OFF/

ZERO 

 

UNIT 

Normal weighing mode     

(more than 3s) 

Enter setup mode  

ON/OFF/

ZERO   

TARE 

Normal weighing mode     

(more than 3s) 

Enter calibration mode when  

 

.Parameters setting summary: SEE SPECS

 

 

Para-

meter

 

x/xy

 

Factory 

Set 

Setting

 

P1.xy

 

00-15

 

05 

Auto-off time: no auto-off;                                          
01-15 minutes auto-off time 

 

P2.x

 

0, 1, 2

 

0=only Hold function; 1=only Print function; 
2=Print and Hold function, 

 

P3.xy

 

0,1,2, 

3-50

 

0=no hold function; 1=hold lager weight reading; 
2=auto release hold function when weight is below 
10d and auto-hold new stable weight (more than 
10d); 3-50=unchangeable reading when the variety 
is 3~50d

 

27 

Содержание PS-500

Страница 1: ...PS 500 SCALE Operation Manual Contents subject to change without notice Version 1 0 2006 01 10 Declarations of compliance...

Страница 2: ...iques de la Classe A prescrites dans le R glement sur le brouillage radio lectrique edict par le minist re des Communications du Canada Warnings Electrical installation For your protection all mains 1...

Страница 3: ...e radio computer or other electronic device These devices emit RF and can causeunstablescalereadings Ifyourscaleeverperformspoorly trymovingthescale to a different room or location Avoid using in cond...

Страница 4: ...ed on a scale following are required on the load cell set to be used with this indicator 1 Sensitivity 0 3mV V 3mV V must be fit to 0 2uV display grad 2 Input Resistor 80 3 Output Resistor 10 K LOADCE...

Страница 5: ...ng range P13 set the scale will show 0 zero point is over the setting range or 0_ _ _ _ zero is below the setting range 3 2 In normal weighing mode and hold down time longer than 3s 1 2 3 Same meaning...

Страница 6: ...will display CAP that means the following data is the full capacity according to your setting of display resolution P7 display division value P8 location of decimal point dot in calibration unit P9 an...

Страница 7: ...nd return back to step 5 for re calibration 8 When CAL P0 is shown again that means the scale will calibrate scale s zero point again Now you can press ON OFF ZERO to exit the calibration mode or Move...

Страница 8: ...splay 1 P A B Item A parameters and one digit can be input 2 P A BC Item A parameters and two digits can be input 3 PAB C Item AB parameters and one digit can be input 4 PAB CD Item AB parameters and...

Страница 9: ...ut format is same as x 1 X 6 Whenthe scaleis stable it willoutputthe currentgross weight tareweight and net weight data including unit automatically one time and not accept data The output format is s...

Страница 10: ...lb Not available 0 01 lb 0 005 kg 0 01 lb 0 2 oz 0 1 lb 0 05 kg 0 1 lb 2 oz 15 1 lb 0 5 kg 1 lb Not available 10 lb 5 kg 10 lb Not available 0 0002 lb 0 0001 kg 0 0002 lb Not available 0 002 lb 0 001...

Страница 11: ...signal is NOT within the power on zero point range x 0 Choose current weight as current power on zero point x 1 Choose calibration zero point as power on zero point x 3 Choose power off zero point as...

Страница 12: ...t data is filled by _ If the zero point is error it will return nine _ characters b The character will be for negative weight or a space character for positive weight Minus sign follows after the firs...

Страница 13: ...urn off the scale 8 Command all others Response Unrecognized command LF CR ETX 2 5 Output status bit meaning Table5 The status bits definition Bit Byte 1 H1 Byte 2 H2 Byte 3 H3 0 stable 0 not under ca...

Страница 14: ...80 000 800 00 8000 0 80000 800000 100000 10 0000 100 000 1000 00 10000 0 10000 0 1000000 Table7 Capacity unit is kg or lb Division set by P8 and P9 P8 1 Resolution set by P7 0 0002 0 002 0 02 0 2 2 2...

Страница 15: ...50 37500 800 0 4000 4 000 40 00 400 0 4000 40000 1000 0 5000 5 000 50 00 500 0 5000 50000 1200 0 6000 6 000 60 00 600 0 6000 60000 1500 0 7500 7 500 75 00 750 0 7500 75000 2000 1 0000 10 000 100 00 10...

Страница 16: ...ignal is too large 6 _ _ _ _ _ weight signal is too small 7 EEP E0 the EEPROM can t be accessed 8 EEP E1 The parameters are not same with backup data 9 EEP E2 The setting parameter s is not in normal...

Страница 17: ...se filtered or un filtered weight A D data ON OFF ZERO Normal weighing mode Zero function or Power off the scale 26 Setup mode Calibration mode Displaying A D code or input voltage mode Exit and back...

Страница 18: ...on zero point 1 FS 1 calibration zero point 2 FS 2 calibration zero point 5 FS 3 calibration zero point 10 FS 4 calibration zero point 20 FS 5 calibration zero point 50 FS 6 calibration zero point 100...

Страница 19: ...6132 Fa x 44 0 870 010 2241 Email sales salterbrecknell co uk Web site www averyweigh tronix com USA Salter Brecknell Weighing Products USA 1000 Armstrong Drive Fairmont MN 56031 Toll Free 800 637 052...

Отзывы: