5. Troubleshooting
41
Error
code
Overview
Displayed
message
Description
Instrument operation
when the error occurs
Solution
E025
Abnormal
Cryo
operation
caution
Cryo Module
Error
T h e c o o l i n g p l a t e
temperature of the Cryo
Module does not reach
the specified value.
Operation continues.
Use the instr ume nt at an
ambient temperature within
10ºC to 35ºC.
Check if the cooling plate has
frost buildup and remove it if
any.
Check the condition of power
connection and power switch
state of the Cryo Module.
I f t h e p r o b l e m p e r s i s t s ,
contact the Sakura instrument
distributor or representative.
Ambient temperature is
too high.
E040
Low battery
caution
Battery voltage
has dropped
The voltage of the clock
battery dropped below
the specified value.
Operation continues.
Contact the Sakura instrument
distributor or representative
and replace the battery within
one month.
E041
Clock reset
The clock was
reset.
Unable to restore time
data. Button battery is
dead.
If the instrument was
operating in the auto
m o d e , o p e r a t i o n
pauses. If the instrument
was operating in the
manual mode, operation
continues.
Contact the Sakura instrument
distributor or representative
and replace the battery.
E042
Invalid clock
The time data is
incorrect
The clock generated an
error. The battery case
came off.
Operation continues.
Check if the date and time
displayed on the screen is
correct. If the date and time
is not correct, set the correct
date and time. If the problem
persists, contact the Sakura
i n s t r u m e n t d i s t r ib u to r o r
representative and replace the
battery.
E050
Momentary
power failure
Momentary
power outage
A m o m e n t a r y p o w e r
failure occurred.
O p e r ation c ontinue s
f r o m t h e O p e r a t i o n
screen. If the light was
o n w h e n t h e p o w e r
was cut off, operation
continues with the light
turned off.
Check the molten state of
paraffin, and then resume
using the instrument.
E100
Abnormal
temperature
sensor
Temperature
sensor error
S i g n a l s c a n n o t b e
r e c e i v e d f r o m t h e
temperature sensor.
Operation aborts.
Turn off the power to the
i n s t r u m e n t a n d c o n t a c t
t h e S a k u r a i n s t r u m e n t
distributor or representative.
E101
E102
A n a b n o r m a l s i g n a l
was received from the
temperature sensor.
Operation aborts.
Turn off the power to the
instrument. When paraffin in
the problem location starts
solidifying (below 60ºC, as
a guide), turn on the power
again. (The power should be
off for at least 10 seconds.)
If th e p r o b l e m p e r s i s t s ,
turn off the power to the
i n s t r u m e n t a n d c o n t a c t
t h e S a k u r a i n s t r u m e n t
distributor or representative.
( C h e c k t o s e e i f a n y
equipment that generates
s t r o n g e l e c t r o m a g n e t i c
waves is nearby.)
Содержание M01-021E-01
Страница 5: ...Table of contents 5 ...
Страница 6: ...1 Introduction ...
Страница 18: ...2 Installation ...
Страница 32: ...3 Explanation of Operations ...
Страница 34: ...4 Error Log ...
Страница 36: ...5 Troubleshooting ...
Страница 46: ...6 Replacement Parts ...
Страница 48: ...7 Glossary of Terms ...
Страница 52: ...52 ...
Страница 53: ...M01 022E 01 ...
Страница 54: ...Please visit our websites www sakura finetek com www sakuraus com www sakura eu ...