TROUBLESHOOTING AND SERVICE
Page 29
5.1
Power Outage
Unlock the Lid Mechanically During
Power Outage
WARNING: MAKE SURE THAT THE ROTATION HAS
STOPPED BEFORE PERFORMING THE WORK.
In the event of a power outage, the instrument lid will
remain locked. The lid can be manually opened, if needed.
1.
Set the power switch to the OFF position.
2.
Remove a small cap located on the left side of the
instrument (Figure 5A)
Figure 5A
3.
Insert a bar for unlocking the lid into the hole (Figure
5B).
Figure 5B
4.
Push in the bar to release the lock mechanism. The lid
can be opened.
CAUTION: Use a rigid bar that is 4 to 6 mm across and
more than 5cm long.
WARNING: BEFORE OPENING THE INSTRUMENT LID
AFTER IT IS UNLOCKED, TAKE INFECTION CONTROL
MEASURES SUCH AS WEARING PROTECTIVE
CLOTHING AND PLACING THE INSTRUMENT IN A
SAFETY CABINET.
5.2 Actions to be Taken When an
Error Occurs
If an error occurs during operation, the instrument
immediately decelerates the motor and sounds the alarm.
After the motor has stopped, the lid lock indicator (LID
LOCK) is turned off. Open the instrument lid and take
corrective actions for the error occurred.
The alarm buzzer can be silenced by pressing any key on the
control panel (
START/STOP, SET
,
UP
and
DOWN
keys).
To exit from the error display, turn the instrument power off
and on again.
Description of the Error Display
(Figure 5C)
Figure 5C
Display of the error code of an error
Display of balance error
Cap
Содержание Cyto-Tek 2500
Страница 1: ...Cyto Tek 2500 Cytocentrifuge Operating Manual ...
Страница 2: ...O p e r a t i n g M a n u a l Cytocentrifuge Cyto Tek 2500 ...
Страница 3: ......
Страница 12: ...1 INTRODUCTION Page 8 ...
Страница 26: ...3 OPERATING INSTRUCTIONS Page 22 ...
Страница 40: ...5 TROUBLESHOOTING AND SERVICE Page 36 ...