Cyto-Tek® 2500 Cytocentrifuge 

QRG | Cytocentrifuge | V-1 

page 3 | 4 

How to load the Guard Bowl 
 

1.

 

Use an even number of specimens chamber 
assemblies.  Arrange  them  in  the  rotor  so 
that  pairs  of  specimen  chamber  holder 
assemblies are opposite pairs. 

Caution: 

Place 

equal 

size specimen chamber holders in opposite  

               

positions. 

 

2.

 

If  there  are  an  odd  number  of  specimen 
samples  to  process,  maintain  the  rotor 
balance  by  adding  one  specimen  chamber 
filled  with  water  to  pair  with  the  last 
specimen. 
 

3.

 

If  all  twelve  specimen  holder  positions  in 
the  rotor  are  not  filled,  additional  pairs  of 
water-filled  speci-men  chamber  holder 
assemblies  may  be  added  as  necessary  to 
keep the rotor balanced. 

 

 

Guard Bowl loading always in pairs 

4.

 

Make  sure  the  cytochambers  are  placed  in 
the right position  
 

 

 

 

Incorrectly set specimen chamber

 

 

Starting the operation 
 

1.

 

Place  the  guard  bowl  into  the  instrument 
and close the lid. 

 

2.

 

Press  the 

up

  or 

down

  arrow  key  until  the 

desired  program  number  is  obtained.  If  a 
desired  program  does  not  exist,  make 
changes to create a new program.  

Note:      

to create a program review the 

Setting Instructions

 in     

               

the Operating Manual. 

 

3.

 

Press  the 

START/STOP

  key  to  start  the 

operation.  The  instrument  is  locked  during 
operation. 
 

 

Starting the operation 

 

 

Содержание Cyto-Tek 2500

Страница 1: ...itro examination of human and animal material for cell structure changes by a pathologist The user instructions for the Cytochambers can be found in the operating manual and in the User Guide V 1 Cytology chambers 1 6 and 12 ml which can be found on the Distribitor area Cyto Tek 2500 Cytocentrifuge Components and Functions 1 Powerswitch Turns the instruments power on and off 2 Control Panel Provid...

Страница 2: ...id Unlock guard bowl Remove guard bowl Control Panel The controls and indicators to display the instrument status ar listed 1 START STOP key Starts and stops the operation 2 SET key Let s the operation mode switch to the Program Mode and Set View Mode 3 down arrow key Decreases a program number and a setting value for speed or proces time 4 up arrow key Increases a program number and a setting val...

Страница 3: ...l twelve specimen holder positions in the rotor are not filled additional pairs of water filled speci men chamber holder assemblies may be added as necessary to keep the rotor balanced Guard Bowl loading always in pairs 4 Make sure the cytochambers are placed in the right position Incorrectly set specimen chamber Starting the operation 1 Place the guard bowl into the instrument and close the lid 2...

Страница 4: ...y and components in an autoclave for 20 minutes at 121 C When no autoclave is available 1 Disassemble the guard bowl unit 2 Soak guard bowl body and components in an ethylalcohol or isopropylalcohol 70 80 3 or sodium hypochlorite 0 5 solution for at 20 minutes 4 Wash with water and soap 5 Rinse with clear water 6 Dry thoroughly when using an oven keep temperature below 120 C Yearly 1 Replace the l...

Отзывы: