gy

 

Quick Reference Guide 

Cyto-Tek® 2500 Cytocentrifuge 

 

page 2 | 4 

 QRG | Cytocentrifuge | V-1 

4.

 

Guard Bowl: 

Accomodates  the  rotor.  The  guard  bowl  is 
detachable, and has an  own detachable lid. 
 

5.

 

Rotor:

 

Is  a  circular  casing  with  compartments 
where up to 12 specimen chambers can be 
palced  at  one  time.  It  also  facilitates  as  a 
stand  when  fluid  samples  are  dispensed 
into the specimen chambers. 
 

6.

 

Bowl Lid:

 

Securely  seals  the  guard  bowl,  a  lock 
button  is  provided  on  top  of  the  lid  knob. 
Locking  the  bowl  lid  by  pushing  down  the 
button and openend by pulling up. 

 

Opening the Cyto-Tek

®

 2500 Cytocentrifuge 

 

 

Guard bowl including rotor 

 

Detachable lid 

 

Unlock guard bowl 

 

Remove guard bowl 

 

Control Panel 

The  controls  and  indicators  to  display  the 
instrument status ar listed. 
 

1.

 

START/STOP key: 

Starts and stops the operation 

 

2.

 

SET key: 

Let’s  the  operation  mode  switch  to  the 
“Program Mode” and “Set & View Mode”. 
 

3.

 

      (down-arrow) key: 

Decreases a program number and a setting 
value for speed or proces time. 

 

4.

 

    

(up-arrow) key: 

Increases a program number and a setting 
value for speed or proces time. 
 

5.

 

Program Number display: 

Displays a program number from 1 to 30. 

 
 
 
 

6.

 

Speed Display: 

Displays a program speed or actual speed. 
 

7.

 

Time display: 

Displays a programmed time or process 
time left. 
 

8.

 

Lid Lock indicator: 

A LED lights up when the lid is locked at 
the start of operation and goes out when 
the lid is unlocked after the motor has 
stopped. 
 

9.

 

Power indicator: 

A LED lights up while the power is supplied 
to the instrument. 
 

 

 

Control Panel 

 

 

Содержание Cyto-Tek 2500

Страница 1: ...itro examination of human and animal material for cell structure changes by a pathologist The user instructions for the Cytochambers can be found in the operating manual and in the User Guide V 1 Cytology chambers 1 6 and 12 ml which can be found on the Distribitor area Cyto Tek 2500 Cytocentrifuge Components and Functions 1 Powerswitch Turns the instruments power on and off 2 Control Panel Provid...

Страница 2: ...id Unlock guard bowl Remove guard bowl Control Panel The controls and indicators to display the instrument status ar listed 1 START STOP key Starts and stops the operation 2 SET key Let s the operation mode switch to the Program Mode and Set View Mode 3 down arrow key Decreases a program number and a setting value for speed or proces time 4 up arrow key Increases a program number and a setting val...

Страница 3: ...l twelve specimen holder positions in the rotor are not filled additional pairs of water filled speci men chamber holder assemblies may be added as necessary to keep the rotor balanced Guard Bowl loading always in pairs 4 Make sure the cytochambers are placed in the right position Incorrectly set specimen chamber Starting the operation 1 Place the guard bowl into the instrument and close the lid 2...

Страница 4: ...y and components in an autoclave for 20 minutes at 121 C When no autoclave is available 1 Disassemble the guard bowl unit 2 Soak guard bowl body and components in an ethylalcohol or isopropylalcohol 70 80 3 or sodium hypochlorite 0 5 solution for at 20 minutes 4 Wash with water and soap 5 Rinse with clear water 6 Dry thoroughly when using an oven keep temperature below 120 C Yearly 1 Replace the l...

Отзывы: