PRINCIPLES OF OPERATION 

  

 
Air is drawn  into the  cupboard  at the  working aperture 

A

 below  the sliding  sash at a velocity 

high  enough  to  ensure  entrainment  of  any  fumes,  odours  or  contaminants  given  off  by  the 
processes carried out within  the  cupboard. 
 
The  fumes  from within  the  cupboard  are drawn  either  upwards  as in the  case of  lighter  than 
air fumes 

to the  extract  slot

  D

, or drawn  across the  work  surface  to the  secondary  extract 

slot 

E

 created by  the back baffle,  as in the  case of heavier  than  air fumes/particulates 

C

 
The  combination of extract  slots 

D

 and 

E

 ensure  that  all fumes/particulates  are carried by the 

moving  body  of  air to  the  extract  slot  and  duct  transition  adaptor 

and  on  into  the  extract 

duct  system

  G. 

 

 

 

 

                                                                  Fig1 

 

Содержание Airone 1200 XP3

Страница 1: ...4 8EE Tel 44 0 1934 421 340 Fax 44 0 1934 641 569 Mailto safelab safelab co uk www safelab co uk Safelab Systems Ltd INSTRUCTION MANUAL XP3 Walk in FUME CUPBOARD Registered No 5336826 England and Wale...

Страница 2: ...instructions Page 7 Location of Fume Cupboard Page 8 Specification Page 9 Specification diagram Page 10 Operating instruction and features Page 11 Operating instruction and features continued Page 12...

Страница 3: ...ended that service and maintenance operations should only be undertaken by SAFELAB Service Engineers or their authorised agents Details of Service Contracts Programmes along with information on these...

Страница 4: ...from escaping while drawing air in through the sash opening at a consistent rate A face velocity the speed at which air is drawn into the cabinet is of importance to the safe and effective operation...

Страница 5: ...cupboard The fumes from within the cupboard are drawn either upwards as in the case of lighter than air fumes B to the extract slot D or drawn across the work surface to the secondary extract slot E c...

Страница 6: ...oard see page 7 for a guide to siting the cupboard Connect the cupboard to the duct system and the control system to the extract fan system before plugging the mains lead The Airone XP3 Ducted Fume Cu...

Страница 7: ...DOCUMENT N M38 ISSUE N 03 ISSUE DATE 26 02 2019 7 GUIDE TO POSITIONING LOCATIONS FOR THE AIRONE XP FUME CUPBOARD...

Страница 8: ...xy powder coated extruded aluminium handle The work tray is grey cast epoxy resin as standard although ceramic work tray are available as an option Sash The vertical sliding sashes on this Walk in Fum...

Страница 9: ...9 SPECIFICATION DIAGRAM Fig 3 1 Hinged top panel 7 Airflow sensor 2 Rear upper baffle 8 Digital controller 3 Light switch 9 Upper sash handle 4 Sash stop 10 Services panel optional 5 Double socket op...

Страница 10: ...ces always consult Safe Systems of Work which should be provided by your employer Air Flow Monitor Controller OPERATING INSTRUCTIONS ENSURE THE UNIT IS PLUGGED IN TO THE MAINS ELECTRICAL SUPPLY 1 Swit...

Страница 11: ...amber light and buzzer will sound and the display will alternate between Sash High and the face velocity to warn that containment levels and airflow velocities are compromised Pressing the ENTER butt...

Страница 12: ...absence of the Fume Cupboard operator If the sash is open with no operator present after a set time default 5 min the alarm will sound and Close Sash will be displayed the alarm will stop sounding whe...

Страница 13: ...DOCUMENT N M38 ISSUE N 03 ISSUE DATE 26 02 2019 13 Wiring diagram Main circuit Controller circuit Inverter circuit where applicable...

Страница 14: ...etergent solution c Inspect the sash mechanism for corrosion and damage lubricate pulley with light machine oil or silicone spray Twelve Monthly Maintenance a Items a to c as above b Check the conditi...

Страница 15: ...the appropriate personal protective equipment PPE Open windows doors fans air conditioning units and movement of people will all cause airflow disturbance which can reduce the effectiveness of the fu...

Страница 16: ...16 DOCUMENT N M38 ISSUE N 03 ISSUE DATE 26 02 2019...

Страница 17: ...DOCUMENT N M38 ISSUE N 03 ISSUE DATE 26 02 2019 17 NOTES left blank for notes...

Отзывы: