20
24 Hour Testing
When activated, provides automatic, unattended testing.
The DisplayMax will scan and record all active channels of the selected channel plan.
The following parameters can be user adjusted either in Setup, see Section 7.0, or directly from the 24 HR Mode.
Parameter Factory
Defaults
Range
24 HR Test
OFF
Current Time
HH:MM:SS
Current Date
MM:DD:YY
First Scan Delay
05 MINS
0 - 59 MINS
00
HRS
0
-
4
HRS
Scan to Scan
00 MINS
0 - 59 MINS
06
HRS
0
-
23
HRS
00
DAYS 0
-
6
DAYS
Number of Scans
04 SCANS
1 - 99 SCANS (Maximum memory capacity is 43 scans)
The data is automatically recorded. Each scan is time and date stamped and named with its scan number.
After each scan, the DisplayMax automatically shuts itself off.
It is then automatically turned on each time a scan is scheduled.
During a 24 HR Scan, all keypad functions are locked out, including the Manual On/Off key.
In between scheduled scans, you may use the DisplayMax in its regular modes.
If the DisplayMax is in operation when a scan is scheduled, a pop-up screen prompts the user to select one of the following options:
Start Scan Now
Delay 5 Minutes
Abort This Scan
Abort Entire 24 Hr. Test
If no option is selected, the scan will begin after a one minute delay.
Scan results can be checked against limits in Recall, they can also be Downloaded and Printed.
DisplayMax Software allows checking and comparing of tests over time.
Important: After enabling the 24 HR Test in the setup menu, you must Press ESC to exit the menu and activate the test. If
the meter is turned off before Escaping the menu, the 24 HR Test will not run.
Make sure there is sufficient battery power or connect the AC Charger before initiating a 24 HR Test.