26
8
Technical Specifications
Table 8: Specifications
Ge
ne
ra
l
Disposable Battery
Chemistry
LiMnO2
voltage
15V
Battery
capacity
5Ah, 200 defibrillator discharges at maximum energy (200J) or 5 hours of
continuous ECG Monitoring
Shelf-life
5 years
Environment
Operating
Temperature
0° to 50° C
Storage
Temperature
-30° to 70° C
Humidity
10 to 95% relative humidity
Altitude
-1,250 to 15,000 ft.; -381 m to 4573 m
Particle and
Water Ingress
IP22
Moth
er Boa
rd
User Interface
Display
Type
LCD with resistive touch
Size
5 inches
Resolution
400*800 pixel
Parameters
Number of Shock, Time, Elapsed Time, Battery capacity
and Menu
ON/OFF Button
Turns Power ON or OFF
Child Selection Button
Sets AED to child mode
Speaker
Provides audio prompts and necessary alarms
Shock Button
It used for semi-automated models (Illuminated button)
HV
B
oa
rd
Waveform
Rectilinear Biphasic
Energy Selections
Automatic pre-programmed selection
Adult mode: 120J, 150J, 200J
Child mode: 50J, 70J, 85J
Charge Time
Less than 15 seconds with new battery pack
Self-Test
Internal self-test includes: charge circuit, shock circuit
Patient Impedance
Measurement
Yes
Impedance Range
25-175 ohms
Shock Delivery
Fully automated model
Semi-Automated model
Communication
UART Communication with Mother Board
UART Communication with ECG Board
EC
G
Bo
ard
ECG Bandwidth
1-30 Hz
Shockable Rhythms
Algorithm spec