Chapter 9: Troubleshooting and Error messages
64
Error messages
Message
Cause
Solution
Remarks
Leads Error Messages
CHECK R
Lead R is not properly
connected to patient.
Make sure that mentioned
electrode
is
properly
connected.
The message is displayed in
red color on the screen.
CHECK L
Lead L is not properly
connected to patient.
Make sure that mentioned
electrode
is
properly
connected.
The message is displayed in
red color on the screen.
CHECK F
Lead F is not properly
connected to patient.
Make sure that mentioned
electrode
is
properly
connected.
The message is displayed in
red color on the screen.
CHECK C1
Improper connection of C1
electrode
Make sure that C1
electrode
is
properly
connected to patient.
The message is displayed in
red color on the screen.
CHECK C2
Improper connection of C2
electrode
Make sure that C2
electrode
is
properly
connected to patient.
The message is displayed in
red color on the screen.
CHECK C3
Improper connection of C3
electrode
Make sure that C3
electrode
is
properly
connected to patient.
The message is displayed in
red color on the screen.
CHECK C4
Improper connection of C4
electrode
Make sure that C4
electrode
is
properly
connected to patient.
The message is displayed in
red color on the screen.
CHECK C5
Improper connection of C5
electrode
Make sure that C5
electrode
is
properly
connected to patient.
The message is displayed in
red color on the screen.
CHECK C6
Improper connection of C6
electrode
Make sure that C6
electrode
is
properly
connected to patient.
The message is displayed in
red color on the screen.
System Messages
Recorder Error Messages
Noisy ECG signal due to AC
power interferences
● Electrodes are placed on
bony site of the patient body.
● Unclean or sulfated
electrodes
● Insufficient gel is applied.
● Contact with metal parts of
bed, trolley, etc.
● Lead wires, patient cable or
power cable fails to make
connection.
● There are other electronic
devices in the vicinity of the
electrocardiograph.
● Improper ambient light for
example using fluorescent
lamp in the room which ECG
record is taken.
● Improper Earth system
● Check electrodes and lead
wires connection.
● Check that lead wires are not
tangled or connected to ground.
● Check that the patient does not
contact the metal parts.
● Check that patient cable and
power cable have no contact.
● Turn on the HUM Filter.
● If the problem persists, unplug
the power cable (the device runs
on the battery).
● If the problem still persists,
noise source may be other
devices, room or its earth system.
Consequently, this room is not
suitable for ECG recording.
Содержание Dena 650
Страница 1: ...User Manual Dena Electrocardiograph DENA 650 Linux AADAT Co D00912 7...
Страница 3: ......
Страница 11: ......
Страница 12: ...Chapter 1 Introduction 1 1 Introduction...
Страница 25: ...Chapter 1 Introduction 14...
Страница 26: ...Chapter 2 System settings 15 2 System settings...
Страница 38: ...Chapter 3 Patient information 27 3 Patient information...
Страница 41: ...Chapter 3 Patient information 30...
Страница 42: ...Chapter 4 Patient preparation 31 4 Patient preparation...
Страница 47: ...Chapter 4 Patient preparation 36 Lead Placement Diagram Figure 4 4 Lead placement digram...
Страница 48: ...Chapter 5 Recording Operation 37 5 Recording Operation...
Страница 52: ...Chapter 6 ECG analysis and measurement 41 6 ECG analysis and measurement...
Страница 58: ...Chapter 7 Data management 47 7 Data management...
Страница 64: ...Chapter 8 Care and Cleaning 53 8 Care and Cleaning...
Страница 71: ...Chapter 8 Care and Cleaning 60...
Страница 72: ...Chapter 9 Troubleshooting and Error messages 61 9 Troubleshooting and Error messages...
Страница 77: ...Chapter 9 Troubleshooting and Error messages 66...
Страница 78: ...Chapter 10 Technical specification 67 10 Technical specification...
Страница 81: ...Chapter 10 Technical specification 70...
Страница 82: ...Appendix 1 Accessories 71 Appendix 1 Accessories...
Страница 84: ...Appendix 2 System parameters 73 Appendix 2 system parameters...
Страница 87: ...Appendix 2 System parameters 76...
Страница 88: ...Appendix 3 Electromagnetic compliance EMC 77 Appendix 3 Electro magnetic compliance...
Страница 93: ...Appendix 3 Electromagnetic compliance EMC 82...
Страница 94: ...Appendix 4 The GLASGOW program 83 Appendix 4 The GLASGOW program...
Страница 101: ......