7
Message
Cause
Solution
Remarks
There’s No
Copy Rec
There is no copy after
turning off & on
If you want to use copy of
last recorded signal, don‟t
turn off the device
The message displays in black
color on the screen.
Measurement Messages
Measurement
Error
Occurrence of any
measurement errors
Find the related error and
do the related solution
The message displays in black
color on the screen
Impossible
measurement
Check Lead
Main leads
disconnection
Check the connection of
leads
The error message prints on
recorder paper
QRS Detection
Error
There are less than 3
QRS complexes
Record again after relaxing
the patient
The error message prints on
recorder paper
Abnormal
Heart Rate
The RR duration is
less than 166 ms
(HR>180 bpm)
|
Record again after check
the patient situation control
The error message prints on
recorder paper
Abnormal
Heart Rate
The RR duration is
more than 750
ms
(HR<40 bpm)
Record again after check
the patient situation
The error message prints on
recorder paper
Weak Signal
(Lead II)
There are more than
50 R-wave because of
noise wave detection
in weak signal with
low amplitude.
Record again after check
the lead position and
patient situation control
The error message prints on
recorder paper
Noisy Signal
There is no R-wave
or they are less than 4
in reference lead(I, II,
V3)
Record again after check
the lead position and
patient situation control
The error message prints on
recorder paper