54
8-3 Power on the electrocardiograph system.
Press Power key to turn on the electrocardiograph.
8-4 Connect the patient cable.
Connect all necessary accessories to patient and the electrocardiograph.
If any sign or error message is observed in the device that may
be due to its failure, please do not use it on the patient.
The battery must be recharged after a while to ensure
adequate electricity reserve. To do so, you only connect the system to
AC INPUT.
The battery needs to be charged after transportation or
storage. If the power cable is not plugged in before turning on the
device, the device may not work properly because of insufficient
power. Connect the device to AC INPUT for about 8 hours while it is
off.
Make sure that the battery indicator lights. If it does not light,
check your local power supply and power cable connection. If the
problem still exists, contact the After -Sales Services department.
The system beeps every time it is powered on. If no beep
sound is heard, the audio system (beep sound, key/touch sound)
may be faulty.