4.1 Physical fundamentals of the power measurement at electro surgery
units
Power output P
L
of electro surgery units is, as it is with all electrical generators, the
product of the voltage U
L
at the load resistor R
L
and the current I
L
by the load resistor
R
L
.
(1)
P
L
= U
L
x I
L
I
L
and U
L
are dependent, as it is shown in the alternate circuit diagram, on the
internal resistor R
i
of the surgical device and from the neutral voltage U
0
from the
HF-generator within the HF surgical device
:
(2)
I
U
R
R
L
i
L
0
(3)
U
I
R
U
R
R
R
L
L
L
i
L
L
0
If you take the right sides of the equations (2) and (3) and put them in equation (1),
then the dependence of power output P
L
from the electro surgery units on internal
resistor Ri and on the load resistor R
L
is recognizable:
(4)
P
U
R
R
R
L
i
L
L
0
2
electrosurgery units
R
i
U
0
I
L
U
L
R
L
Содержание HF-400
Страница 19: ...S P L HF 400 User manual 19 6 Measurement Picture 6 1 Picture 6 2...
Страница 20: ...S P L HF 400 User manual 20 Picture 6 3 Picture 6 4...
Страница 21: ...S P L HF 400 User manual 21 Picture 6 5 Picture 6 6...
Страница 22: ...S P L HF 400 User manual 22 Picture 6 7 Picture 6 8...
Страница 23: ...S P L HF 400 User manual 23 Picture 6 9 Picture 6 10...
Страница 24: ...S P L HF 400 User manual 24 Picture 6 11 Picture 6 12...
Страница 25: ...S P L HF 400 User manual 25 Picture 6 13 Picture 6 14...
Страница 26: ...S P L HF 400 User manual 26 Picture 6 15...