Page 9
Charcoal Air Filter Replacement
The charcoal air filter element, located on the rear of the unit, should be replaced after every
25 flush and purge services. It may be necessary to replace the element sooner if the operator
notices a stronger than usual odor emitted from the unit during the purge process.
1.
Turn the wing nut on the filter counterclockwise and remove the retaining plate.
2.
Remove the old filter element and replace with a new one.
3.
Replace retaining plate and tighten wingnut to finger tight only.
Coalescing Filter Element Replacement
The coalescing filter element must be replaced after every 100 flush and purge services. In
some cases the element may need to be replaced sooner if the shop air supply contains high
amounts of moisture and oil. A clogged element will restrict the airflow in the purge process
and should be replaced immediately.
1.
Turn counterclockwise on the filter bowl to remove it.
2.
Remove the filter element by turning it counterclockwise.
3.
Replace the used element with a new one. Hand tighten only.
4.
Replace the filter bowl by turning it clockwise.
Содержание ACF-3000
Страница 13: ...Page 12 Flow Diagram ...