drill may be used only to remove root canal filling partially, and the finishing drills may be used only to pre-
pare root canal.
D.T. drills must be sterilized individually in an autoclave with the following settings**:
•
Wrap individually in disposable sterilization pouches or tubing that comply with the regulation standard
in force in the country of use (e.g EN ISO 11607-1),
•
Autoclave: type B complying with regulation standard in force (e.g EN 13060),
•
Sterilization temperature: 134°C - sterilization time: 18 minutes.
Use a new drill after 12 -15 uses.
For profoundly flared or ovoid canals, the use of accessory post, such as Fibercone
™
, is indicated for best
results.
STORAGE
Store the sterilized components in a dry, dust-free place.
If the packaging’s integrity appears to be compromised, before using again, place the component in a new
pouch ad re-sterilize according to the protocol defined before.
SAFETY AND LIABILITY
RTD declines any liability or compensation for possible damage due to:
-
use of instruments which do not belong to the system and which may impair the function,
-
failure to observe the instructions for use.
The user is responsible for testing the material for its suitability and use for any purpose not explicitly stated
in the instructions for use. Selection of the correct drill and post is the responsibility of the user.
For dental use only.
*
Refer to the manufacturer’s instructions for use.
**
Please refer to your autoclave manufacturer for specific sterilization instructions if needed