VW2106 Vibrating Wire Readout Manual
RST Instruments Ltd.
Page 16
F
IGURE
4-5
D
ATA LOGGING INTERVAL
5.
Using the up/down arrows, select the data logging number and press
Enter
. For
example, if the number 4 is selected, the data will log 4 times every interval, as
illustrated in Figure 4-6. To continuously log data until the memory is full, set the
data logging number to 0.
F
IGURE
4-6
D
ATA LOGGING NUMBER
6.
The readout will start the data logging process and will end with a message
stating “Logging completed”.
ESC
can be pressed any time to abort the data
logging process.
4.5
D
ATA
L
OGGING
T
HROUGH
S
TORE
D
ATA
Once the intervals and iteration preferences have been defined (see section 4.4 step
4), data logging can also be accessed through the
Store Data
screen.
1.
Using the up/down arrows, scroll to the
Store Data
screen and press
Enter
.
2.
Select the location where the data will be stored. Press
Enter
.
3.
Press and HOLD the
Enter
button to commence data logging. The logging