VW2106 Vibrating Wire Readout Manual
RST Instruments Ltd.
Page 15
F
IGURE
4-4
S
TORING A READING WITH
6
SENSORS
NOTE:
A
VERAGE READINGS WILL NOT BE STORED IN THE UNIT
’
S MEMORY
.
4.3
R
EVIEWING
D
ATA
The following instructions outline the steps to review readings on the VW2106.
1.
Press any key to turn the readout on.
2.
Using the up/down arrows, scroll to the
Memory
screen and press
Enter.
3.
Using the up/down arrows, scroll to
Review Data
and press
Enter.
4.
Scroll to the desired location using the up/down arrows and press
Enter
.
5.
If a location contains more than one reading, the readings can be scrolled
through using the up/down arrow.
4.4
D
ATA
L
OGGING
The VW2106 has a basic datalogging function where the user can set the
datalogging interval and the number of iterations.
The following instructions detail the steps to log data at a given location.
1.
Use the up/down arrows to scroll to the
Memory
screen and press
Enter
.
2.
Use the up/down arrows to scroll to the
Data Logging
screen and press
Enter.
3.
Use the up/down arrows to select the location to record the data. If no locations
are currently defined, the
readout with respond with “No Locations.” See section
3.5 for detail about how to define locations.
4.
Use the up/down arrows to set the reading interval and press
Enter
. For
example, if 4 is selected, the data will log every 4 seconds (Figure 4-5).