![RST DLS-35 - 10 Series Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/rst/dls-35-10-series/dls-35-10-series_instruction-manual_1496809010.webp)
DLS–35
©
Dinel, s.r.o.
10
h
In case of
vertical installation
for sensing the level
of
unknown (conductive and non-conductive)
liquids
in tanks or sumps, it is appropriate to bend
the electrode into a right angle. This will increase
the local sensitivity and accuracy of sensing the
level at the spot of the bend.
For: DLS–35_–30
h = 50 to 200 mm
Figure 7: Bending of the electrode end to sense
unknown fluids
When weather conditions (wind, rain, snow) clearly
influence the electrode (open sumps), we recom
-
mend using types with an insulated electrode
(DLS-35_-21,22,24,25,31).
mm
200
50
h
E
3
2
m
E
4
3
p
E
3
4
s
E
3
1
E
10
1
a
20
E
40
b
50
E
10
c
40
E
40
d
20
E
20
k
Mounting in a
bypass measuring tube
. We recommend upholding the tube diameter.
For types: DLS–35_–20,21,22,23,24,25, 30, 31
E– Electrode length in mm
Figure 8: Bypass measuring tube