2
3
Function
Maximum Input
Voltage AC or DC
1000V AC RMS/1000V DC
Low Z
600V AC RMS/600V DC
μA, mA Current AC/DC
800mA 1000V fast acting fuse
10A Current AC or DC
10A 1000V fast acting fuse (10A for 30 seconds
max. every 15 minutes)
Resistance, Continuity, Diode Test,
Capacitance, Frequency, Duty Cycle 600V AC RMS/600V DC
Temperature
600V AC RMS/600V DC
Category Rating Brief Description
Typical Applications
CAT II
Single phase
receptacles
and connected loads
-Household appliances, power tools
-Outlets more than 30ft(10m) from a CAT
III source
-Outlets more than 60ft(20m) from a CAT
IV source
CAT III
Three phase circuits
and single phase
lighting circuits in
commercial buildings
-Equipment in fixed installations such
as 3-phase motors, switchgear and
distribution panels
-Lighting circuits in commercial buildings
-Feeder lines in industrial plants
-Any device or branch circuit that is close
to a CAT III source
CAT IV
Connection point
to utility power and
outdoor conductors
-Primary distribution panels
-Overhead or underground lines to
detached buildings
-Incoming service entrance from utility
-Outdoor pumps
Insulation
Class 2, double insulated
Enclosure
Double Molded, IP67 waterproof and dustproof
Diode Test
Test current 1.5mA typical, open circuit voltage 3V typical
Continuity Test
Audible signal if the resistance is approx. 50 ±5 or less
Low Battery Indication
" " is displayed
Battery life
150 hours(light off)
Display
6000 count LCD display
Over Range Indication
“OL” is displayed
Polarity
Minus symbol “-” is displayed for negative polarity
Measurement Rate./
3 readings per second, nominal
Auto Power Off
After approx. 15 minutes of inactivity
Input Impedance
10M AC/DC Voltage
Low Z
Approx. 3k nput impedance
AC Response
True RMS
AC Bandwidth
2kHz
Batteries
Four “AAA” 1.5V alkline batteries 800mA 1000V (6.3 x
32mm) fast blow
Fuses
10A 1000V(10 x 38mm) fast blow
Operating Environment
32°F to 104°F(0°C to 40°C) at <70% relative humidity
Storage Environment
14°F to 140°F(-10°C to 60°C) at <80% relative humidity
2000
Input Limits
3. General Specifications
Operating Altitude
meters maximum
Dimensions/Weight
170 x 75 x 48mm/ 386g
Safety
Complies with UL 61010-1 v.3 for measurement Category lV
600V and Category lll 1000V, Pollution Degree 2
3-1. International Safety Symbols
3-2. Safety Category Ratings
Potential danger. Indicates the user must refer to the manual for important safety
information.
Indicates hazardous voltages may be present.
Equipment is protected by double or reinforced insulation.
Indicates the terminal(s) marked must not be connected to a circuit where the voltage
with respect to earth ground exceeds the maximum safety rating of the meter.
True RMS Digital Multimeter/ English
True RMS Digital Multimeter/ English
13/12/2017 Version No. 001