![RS PRO RS-136 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/rs-pro/rs-136/rs-136_instruction-manual_1495969037.webp)
Temperature Calibrator (RTD + TC) / EN
_________________________________________________________________________________
03/27/2020 Version No. 1 34
RTD source in
℉
Accuracy is based upon 4W connection, driving voltage is less than 1.7V and the
excitation current is based upon 0.5mA to 5mA (0 to 400Ω) and 0.05mA to 5mA
(400Ωto 7000Ω). For 3-wire RTD source, assuming all three RTD leads are
matched, add 1.8
℉
(Pt10 and Cu10), 1.1
℉
(Pt50 and Cu50), 0.7
℉
(Other
RTD types) to the specifications .
RTD Type
(α)
Source (°F)
Range
Accuracy
10
Ω
Pt(385)
-328 to 212
2.7
212 to 1472
3.24
50Ω
Pt(385)
-328 to 212
0.72
212 to 1472
0.9
100Ω
Pt(385)
-328 to 212
0.36
212 to 1472
0.015%+0.324
200Ω
Pt(385)
-328 to 212
0.36
212 to 1166
0.015%+0.324
500Ω
Pt(385)
-328 to 212
0.54
212 to 1166
0.015%+0.504
1000Ω
Pt(385)
-328 to 212
0.36
212 to 1166
0.015%+0.324
100Ω
Pt(3902)
-328 to 212
0.36
212 to 932
0.015%+0.324
100Ω
Pt(3916)
-328 to 212
0.36
212 to 1166
0.015%+0.324
100Ω
Pt(3926)
-328 to 212
0.36
212 to 1166
0.015%+0.324
10Ω Cu(427)
-148 to 500
2.7
120Ω Ni(672)
-112 to 500
0.27
50Ω Cu(427)
-292 to 392
0.72
100Ω Cu(427)
-292 to 392
0.36
YSI400
59 to 122
0.36
Temperature coefficient
:
(± 0.002% of reading ±0.002% of range)/°C (<18°C or
>28°C)