Temperature Calibrator (RTD + TC) / EN
_________________________________________________________________________________
03/27/2020 Version No. 1 18
3.6 DC Current calibration using OHM measurement setup (4, 3,
and 2W)
The calibrator can also be used to calibrate 4 values of current (100
μ
A, 250
μ
A, 1mA, and 2mA). Users can enter the mode of OHM measurement and
press button 5 to enter manual mode with MAN displayed on LCD. Connect
the instrument to be calibrated. Press the button 5 repeatedly to circulate
through 4 fixed values (100
μ
A, 250
μ
A, 1mA, and 2mA). The accuracy of
these values of current is
±
0.015%
±
0.05
μ
A.
Note: The value of DC current is displayed at the upper right corner of LCD
next to the MAN symbol.