RS PRO IRT 1900 Скачать руководство пользователя страница 6

 

EN-4 

IV.  PARTS & CONTROLS 

4-1 Name of Parts and Positions:  

TEST

180°

X1/2

X1
X5

1

2

3

4

5

6

7

9

8

FI

K

EUR

  

 

(1) LCD

 

(2) Proceed test button. 
(3) Phase angle select button. 
(4) x 1/2, x 1, x 5 test current factor select button. 
(5) x 1/2, x 1, x 5 test current factor select button. 
(6) Trip current settings 
(7) Wiring indication. 
(8) Storage compartment 
(9) AC voltage input. 
(10) Earth probe socket. 
(11) Top cover. 
(12) Test lead (AC cord) 
(13) External earth probe. 

Содержание IRT 1900

Страница 1: ...Instruction Manual IRT 1900 Digital RCD Tester FR EN IT DE ES TEST 0 180 X1 2 X1 X5...

Страница 2: ...f Parts and Positions 4 4 2 LCD Display 5 4 3 Connections For RCD Testing 6 V OPERATING INSTRUCTIONS 6 5 1 Warning initial checks to be carried out before any testing 6 5 2 Preparation 7 5 3 Testing 8...

Страница 3: ...tions rated at 230V AC 10 Under No circumstances must it be connected to higher voltage or phase to phase Never open the instrument case There are no user serviceable parts inside If the unit does not...

Страница 4: ...Resilient case protects against everyday knocks 0 and 180 degree phase angle switch permits quick tests and consistent readings III SPECIFICATION Test current range 10 20 30 100 300 500 mA Test curren...

Страница 5: ...IEC Test Lead External Earth Probe Environmental Conditions The product is designed for indoor use and up to a maximum altitude of 2000m Installation Category 300V CAT III single phase 230V Pollution...

Страница 6: ...roceed test button 3 Phase angle select button 4 x 1 2 x 1 x 5 test current factor select button 5 x 1 2 x 1 x 5 test current factor select button 6 Trip current settings 7 Wiring indication 8 Storage...

Страница 7: ...ion 3 x 1 2 test current factor indication 4 x 1 test current factor indication 5 x 5 test current factor indication 6 0 180 phase angle indication 7 180 360 angle indication 8 Analog indication 9 Dis...

Страница 8: ...0 100 P N V OPERATING INSTRUCTIONS To ensure safe and proper use of this instrument the following instruction must be followed 5 1 Warning initial checks to be carried out before any testing a Always...

Страница 9: ...rectified before proceeding c If other loads are connected to the test circuit these may cause erroneous readings To make the most accurate readings loads on the circuit should be disconnected d Use o...

Страница 10: ...nctioning correctly and OL will be displayed d While the test is being conducted the units displayed will be ms for milliseconds e If the Press to Test button is released the result will be displayed...

Страница 11: ...tch again Press the Press to Test button the breaker should trip e The readings obtained in 5 4 d and 5 4 c should both be within the trip time specified for the breaker at its rated tripping current...

Страница 12: ...tch to the 180 position and press the Test button g The trip times of the last two measurements should both be less than 40 ms If the tripping times are greater than 40 ms the RCD may be faulty and mu...

Отзывы: