EN - 12
Range
DC Accuracy
AC Accuracy
±(
1.0% + 2dgt)
±(
1.5% + 5dgt)
50Hz ~ 500Hz
200.0V
1000V
1-3 Electrical Specifications
Accuracy is
±
(% r number of
digits) at 23°C (73.4°F)
±
5°C (
±
9°F)
< 80%RH.
Temperature coefficient :
Add 0.2 x (Specified accuracy) / °C (1.8°F),
< 18°C > 28°C (< 64.4°F >82.4°F).
DC / AC Volts
Over voltage protection :
DC 1000V / AC 750V
Input Impedance :
2 M
Ω
// less than 100pF.
CMRR / NMRR:
(Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
AC
: CMRR > 60dB at DC, 50Hz / 60Hz
V
DC
: CMRR > 100dB at DC, 50Hz / 60Hz
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type:
Average sensing rms indicating.
AC Current
Function
Range
Accuracy
A
(50~60Hz)
0.0 ~ 200.0A
±(
3.0% + 3 dgt)
Overload protection :
400 Arms
AC Conversion Type :
Average sensing rms indication.
* Adjacent conductor influence :
<0.08 A/A