13/08/2017 Version No. 001
MeterScope/English
MeterScope/English
13/08/2017 Version No. 001
27
26
Fuses are FF 0
.
8A 1000V
+
FF 10A 1000V
5°C to 40°C (41°F to 104°F)
-20°C to 60°C (-4°F to 140°F)
Max 80% up to 31°C (87°F) decreasing linearly
to 50% at 40°C (104°F)
<80%
7000ft. (2000meters) maximum.
This meter is intended for origin of installation use
and protected, against the users, by double
insulation per EN61010-1:2010 and
EN61010-2-030:2010, EN61010-2-033:2012 to
Category IV 600V and Category III 1000V
/
/
Fuses
Operating Temperatur
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Safety
Bluetooth specification
Built-in lithium
Li-ion Battery
Version 3.0+EDR,Frequency range 2400 MHz …
2483.5 MHz.(ISM-Band),Guard band 2 MHz < F
< 3.5 MHz.Modulation method GFSK,1 Mbps,0.5
Gaussian; Receiving signal range -82 to -20 dBm;
Transmission power Minimum: -18dBm to +4 dBm
Diameter: 20.0mm; High 3.2mm; Typical Weight
3.0 grams (0.10 oz.); Designation: ANSI / NEDA-
5004LC,IEC-CR2032; Normal Voltage: 3.0 Volts;
Typical Capacity: 240 mAh ; Storage 5 Year
Chemical type: Lithium polymer,Standard: GB/T
18287-2000; Normal Voltage: 7.4 Volts; Charge
up Voltage: 8.4 Volts; Typical Capacity: 2400 mAh.
Cycle life: 500 times ;
7.Specifications
Function
DC Voltage
Range
[1]
50mV
[1]
500mV
5V
50V
500V
1000V
Resolution
0.001mV
0.01mV
0.0001V
0.001V
0.01V
0.1V
Accuracy
(0.05% + 20)
(0.025% + 5digits)
(0.025% + 5digits)
(0.025% + 5digits)
(0.05% + 5digits)
(0.1% + 5)
Function
AC Voltage
Range
500mV
5V
50V
500V
1000V
All AC voltage ranges are specified from 5% of range to
100% of range
50mV
Resolution
0.01mV
0.0001V
0.001V
0.01V
0.1V
0.001mV
Accuracy
50 to 10000Hz
50/60Hz(0.3% + 25)
<1KHz(0.5% + 25)
<5KHz(3% + 25)
Function
(AC+DC)
Range
500mV
5V
50V
500V
1000V
50mV
Resolution
0.01mV
[1]
0.0001V
0.001V
0.01V
0.1V
0.001mV
Accuracy
0 to 1000Hz
<1KHZ(1% + 25)
<10KHZ(3.5% + 25)
[1] When using the relative mode (REL Q) to compensate for offsets.
[1] Add 1% above 5k