15.8.2 - Pipette looks like zebra crossing
If this always occurs in the same pipette, check the bottom of the pipette for the
following:
1. Glass may be chipped.
▪ Replace pipette.
2. Dirt, e.g. dried blood.
▪ Clean the pipette.
▪ Check disinfectant flow at the rinse nozzle.
3. Perpendicularity and straightness of the bottom face.
▪ Replace pipette.
If this happens randomly or with each pipette, check the following:
1. Fill nozzle O-ring or flat washer.
2. Fill nozzle alignment to pipette.
▪ Check the nozzle arm is tight on the rear vertical shaft. Usually engineer’s
assistance is required.
It also could be that there was not enough blood in the sample tube left for complete
filling.
A pipette which looks like zebra crossing gives ESR Error 3.
15.8.3 - One air bubble about 5 mm under meniscus
The filling (aspiration) speed is not critical but should be within certain limits.
1. If just one air bubble is found about 5mm below the meniscus, the filling speed
may be too high.
2. The blood column should not exceed the Fill stop sensor by more than 10mm.
One air bubble can result in ESR Error 3.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
155
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...