ESR Error code messages:
0
No errors
1
No cells/plasma found
Error
No contents could be detected in the pipette.
2
ESR Probably > 140 mm
Error
Extremely high ESR value.
3
Too many borders found
Error
More than three borders found, possibly air
bubbles. See Section Troubleshooting
.
4
Column height <nnn>
Warning
Column height must be between 180 and
210mm. <nnn> = the actual column height.
5
Measure error
Warning
The down count is not equal to the up count
from the measure head.
6
Bubbles on top
Warning
Air bubbles on top of the ESR. See Section
7
Limit error
Error
One of the following limits are out of the setting
range:
•
ESR Time
•
Column height
•
Dilution
•
Bubbles on top
•
Hazy aspect
•
Temperature
9.2 - Result Printout
The results of the ESR measurements can be send to the printer. The layout of the report depends on the
selection of the 60- or 30 minute method.
Columns:
1. Sample ID=Patient number.
2. Hh=Not corrected 30-minute
result (only in use if 30 minute mode is active).
3. ESR=Not corrected 60-minute
result.
4. Tc=60-minute ESR result in millimeters, corrected for
18
°C. (only in use if temperature correction is
active).
5. Aspect (clear, hazy (see page 254)).
6. Pip.=Sedimentation pipette number (number on the pipette belt).
Reporting
MRN-071-2-EN version 7.02 Starrsed RS
117
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...