Select a communication port to communicate with the ESR analyser unit.
Communication with LIMS can be a serial or an ethernet connection.
Use Serial output
:
1. Serial output comport COM2 (Default)
2. Baud rate. Selectable 1200, 4800 and 9600 *
3. Data bits. Selectable 7 or 8 data bits
4. Parity bit: Selectable none, space, mark, even and odd
5. Stop bits: Selectable 1, 1.5 or 2 stop bits
6. Flow control: Selectable for;
▪ None
▪ XON/XOFF
▪ RTS/CTS
▪ DTR/DSR
▪ XON/XOFF & RTS/CTS
▪ XON/XOFF & DTR/DSR
*When the selection button is pressed, the virtual keyboard pops up. Type the correct numbers into the
numerical fields, for instance in the baud rate field 9600.
Use TCP/IP
: Select IP Address and TCP Port as supplied locally.
6.7.2.1 - Set protocol settings
The following protocols can be selected for out putting data;
1. No Serial output.
2. MECHATRONICS-01 Bidirectional
3. MECHATRONICS-02 Unidirectional
4. SE 9000.
5. SE 9000 Unidirectional
6. R-3500.
7. R-3500 unidirectional.
8. R-3500 EPU
9. Compact unidirectional
When selected the following keys will pop-up.
▪ Checksum On/Off.
▪ 30 Minute method On/Off.
▪ Ack/Nack On/Off.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
101
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...