Connect the plug from the power adaptor to
the power input port on the CO Monitor.
Then connect power adaptor or battery
clips to power source.
Fig. 1.8
Fig. 1.9
The air inlet is fitted with a filter which should
be removed and cleaned on a regular basis.
Filter
AIR CONNECTIONS (cont.)
POWER CONNECTIONS
7
If you are using an external alarm device,
connect the plug of the device to the
external power outlet. The external alarms
should be located close to the work area so
they can be seen and heard by operators.
Fig. 2.0
POWER CONNECTIONS (cont.)
If you choose to organize your own external
alarm device you are recommended to use
a qualified Electrician. To actuate external
devices with a relay, the relay coil should be
connected between pin 3 main power) and
pin 4 (alarm). The relay should be a 12v
DC, low power type.
EXTERNAL ALARM WIRING
DIAGRAM
Fig. 2.1
Amphenol C016 30C006 100 12
5.
Black
- Ground
4.
Orange
- Mains supply
+ 12 v power
1.
Blue
- Constant Supply
Alarm, active low
2.
Red
- Constant Supply
Power Output
3.
Purple
- Mains Supply
Alarm, active low
6.
Black
- Ground