14

Saving a Data Record (cont.)

The “Save Log Options” window will display 
and you have the option to choose 
between saving the entire log or alarms 
only, (alarms only will also display when an 
alarm has been cancelled). You also have 
the option to customize the information by 
ticking the desired boxes and clicking on 
“Save Selected Records”.

Click on your choice of file then a SAVE AS 
window will display where you can have the 
opportunity to rename the file and save in 
the desired location. The file is saved as a 
.txt file.

To clear the graph, click on Clear. (This will 

only clear the data from the current screen, 

records will still be held on the data logger). 

Fig. 3.2

Fig. 3.3

If you have any questions about the data record you can email the .txt file to 

[email protected]

for one of our team to review and assist with any

questions.

Содержание RADEX 08-200

Страница 1: ...INSTRUCTION MANUAL...

Страница 2: ...tion Maintenance Service Parts Accessories Read all instructions warnings before using this product Keep this manual for future reference RPB Respiratory T 1 866 494 4599 F 1 866 494 4509 E info rpbre...

Страница 3: ...which can be carried or mounted either on a Radex Airline Filter or on a wall All system lights and alarms are located on the outside of the protective case An External power outlet is located on the...

Страница 4: ...when testing the alarm Do not open the monitor panel as the unit is equipped with a tamper switch and it will void all warranties Opening of the panel will result in a record being placed in the trace...

Страница 5: ...n on your start menu and desktop this will give you easy access to the CO monitor data records To retrieve information from the CO monitor you will have to have the PC connected to the monitor using t...

Страница 6: ...complete the registration of the unit for your 12 month warranty Your Distributor may have already registered the Unit for you this can be checked by clicking on Show Registration BEFORE YOU USE YOUR...

Страница 7: ...hen turn clockwise The case will click into position 6 AIR CONNECTIONS Then connect the air supply hose to a Grade D air source which provides between 10 125psig Fig 1 6 Fig 1 7 Connect the air supply...

Страница 8: ...uld be located close to the work area so they can be seen and heard by operators Fig 2 0 POWER CONNECTIONS cont If you choose to organize your own external alarm device you are recommended to use a qu...

Страница 9: ...r and software version then it performs a self diagnostic test Should there be a fault during the self diagnostic test the screen will display a phone number and email address for you to contact a ser...

Страница 10: ...already sounding the patterns are inverted SELF TEST FAILURE The unit goes through the self test and shows the success or failure of each test If certain parts of the self test fail the unit will not...

Страница 11: ...and onboard battery charger which is operational when the unit is connected to a power supply The battery is for back up only and is used in the event of a power failure The model 08 200 is not desig...

Страница 12: ...application may not be compatible with all types of PC operating systems PC SET UP A USB cable is provided for connecting the Monitor to a PC The first time it is connected to a PC a message will pop...

Страница 13: ...th trouble shooting to find out the cause of an air quality issue like vehicles parking beside air intakes or compressors overheating Fig 2 6 To stop the real time monitoring click on the REAL TIME OF...

Страница 14: ...the OPEN COM button at the bottom right of the main window Fig 2 4 Start and end dates and times can be set to download a particular section of the Data record Fig 2 8 Once you have set your start and...

Страница 15: ...Selected Records Click on your choice of file then a SAVE AS window will display where you can have the opportunity to rename the file and save in the desired location The file is saved as a txt file...

Страница 16: ...acy on a monthly basis this can be done using the Zero Air and 20ppm CO calibration kit p n 08 228 Then press the OK button to begin the Calibration procedure You will be required to press OK between...

Страница 17: ...l operation If there has been an error during the calibration the screen will display an error message You will be provided with a phone number and email address to contact the factory for servicing s...

Страница 18: ...and the sensor can only be replaced by an authorized service center as the sensor has to be set to the factory calibration in order to provide accurate gas detection MAINTAINENCE RECORD T 1 866 494 4...

Страница 19: ...Calibration Kit c w Zero Air 20ppm Carbon Monoxide Cylinders Replacement Zero Air 20ppm Carbon Monoxide Cylinders Aluminium Wall Bracket Mounting Adaptor and Screws Strobe light c w 50ft Cable Externa...

Страница 20: ...y standards worldwide This is the lightweight robust helmet you have been waiting for It helps to reduce fatigue reduce down time and increase productivity These large capacity six stage Filter Cartri...

Отзывы: