background image

RMS-LOG Wireless / LAN Data Loggers 

 

E-M-RMS-LOG-V1.4.docx 

Instruction Manual 

 

© 2018 

 ROTRONIC AG 

 Bassersdorf 

 Switzerland 

 

Page 31 of 35

 

Battery (No battery option with the RMS-D-L). 

Polarity protection 

Yes 

Current consumption 

<100 mA 

Battery life 

3 years at 23 °C and interval of 1 minute) 

 

Start Time and Measurement Interval 

Start time 

LAN logger: 10 s (typical) 

Wireless logger: 1 s (typical) 

LAN display: <20s 

Measurement interval 

10 s to 15 min 

Refresh rate (RMS-D-L) 

10 s 

 

Housing Specifications 

Housing material 

ABS 

Dimensions 

105 x 113 x 38 mm 

Weight 

200 g 

 

Conformity 

EMC directives 

RMS-LOG-L 

EMC Directive: 2014/30/EU 

LVD Directive: 2014/35/EU 

 

EN 61326-1:2013 (Industrielle Umgebung) 

IEC 61326-1:2012 ed2.0 (Industrial Environment) 

EN 55011:2016, class B 

IEC CISPR 11:2015; class B 

EN 55032:2016 

IEC CISPR 32:2016 

EN 61010-1:2010 

IEC 61010-1:2010 

EN 50581:2012 

IEC 50581:2013-02  

 

Performance criterion: www.rotronic.com 

Содержание RMS Wall mount Data Logger

Страница 1: ... 2018 ROTRONIC AG Bassersdorf Switzerland RMS Wall mount Data Logger ...

Страница 2: ... 9 Indicator and Button 8 1 10 Interface 9 1 11 MODBUS communication protocol 9 2 Dimensions 12 3 Installation 13 3 1 Drilling Template Wall Bracket 19 4 Electrical Connections 20 4 1 Battery 20 5 Operation 21 5 1 Default Configuration 21 5 2 Configuration of the LAN Devices with RMS CONFIG 21 5 3 General Manipulations 22 5 4 Integration in the RMS WEB Software 23 5 5 Function Overview 26 6 Mainte...

Страница 3: ...iew The Rotronic Monitoring System RMS is a network comprising various devices and the RMS server software The software is the heart of the system It collects all measured data of the devices and saves it in the database The individual devices work as input modules data loggers and as output modules displays analog outputs switched outputs The user can view the system data at any time on a PC lapt...

Страница 4: ...rovides two analog voltage or current outputs or is also available as variant with two solid state relays in order for example to switch alarm lamps Input Module Records voltage or current signals from analog devices such as particle counters flow transmitters or CO2 probes For example HF5 transmitter humidity temperature AF1 transmitter air flow CO2 transmitter CO2 PF4 transmitter differential pr...

Страница 5: ...f up to 44 000 pairs of measured values o Transfer of the recorded data to the RMS software o Calibration and adjustment of the probe o Firmware update 1 4 RMS Display Order code RMS D L The RMS Display is a freely configurable device The unit can be placed anywhere not necessarily next to the measurement point The display allows the visualisation of measurement values conditions and alarms of dev...

Страница 6: ...not battery powered 24 VDC 10 100 mA2 via terminals V V Power over Ethernet PoE per standard IEEE 802 3af Class 1 Note on the batteries The AA batteries are lithium thionyl chloride batteries available in the industrial trade All RMS input modules are designed for this type of battery Only batteries of the same type or with identical characteristic values may be used as replacement batteries 1 5 1...

Страница 7: ... 868 V Data logger 1 x analog input 868 0 10V RMS MLOG B xxx Mini data logger integrated temperature humidity probe rh C 868 915 MHz RMS MLOG T xxx Mini data logger integrated temperature probe C 868 915 MHz RMS MLOG T10 xxx Mini data logger external NTC probe C 868 915 MHz NTC needs to be ordered separately RMS MLOG LGT 868 Mini data logger integrated light sensor 868 MHz RMS MDI 868 Mini data lo...

Страница 8: ...used to start the device or switch it off in battery mode The LED indicates the device status and whether it was possible in the current measurement interval to carry out a valid measurement and send the data to the monitoring system LED Status Indicator for LAN Devices Pairing Trigger Action LED 1s press Confirms pairing n x orange the LED flashes orange when the pairing demand is open Remove pai...

Страница 9: ...s deleted Device status update Trigger Action LED 1s press Shows the current status Green shows the ISM connection see below 1 x red battery low 2 x red there is no connection to the server Automatic every 5 seconds Shows the current status 1 x green the connection to the server is good 2 x red there is no connection to the server Wireless range Trigger Action LED 1s press Shows the current wirele...

Страница 10: ...ld be blocked 1 11 1 Function 04 Read Input Register Device Data Register Parameter Data type 30001 Serial number Unsigned 32 Bit 30002 30003 Serial number oft he sensors Unsigned 32 Bit 30004 Float Values Register Parameter Data type 31001 Value 1 Measurement value sensor 1 Float 32 Bit 31002 31003 Value 2 Measurement value sensor 2 Float 32 Bit 31004 Integer Values The measurement values are onl...

Страница 11: ...1002 0 0x0000 31003 17530 0x447A 31004 0 0x0000 Word Swap Mixed Endian 31001 0 0x0000 31002 16948 0x4234 31003 0 0x0000 31004 17530 0x447A Byte Swap Middle Endian 31001 13378 0x3442 31002 0 0x0000 31003 31300 0x7A44 31004 0 0x0000 Byte and Word Swap Little Endian 31001 0 0x0000 31002 13378 0x3442 31003 0 0x0000 31004 31300 0x7A44 Software Compatibility The logger is designed for use with the RMS s...

Страница 12: ...ONIC AG Bassersdorf Switzerland Page 12 of 35 2 Dimensions The RMS data logger the gateway and the display are all integrated in the same wall housing All dimensions are shown in Figure 2 Figure 2 Dimensions of the wall housing Figure 3 Mounting bracket with dimensions for drill holes ...

Страница 13: ...scribes installation of the wall housing Installation of the DIN top hat rail housing requires no special handling All terminal markings are identical to those of the wall housing The following instructions describe installation of the data logger step by step 1 Press a blunt object e g screwdriver in the AC1321 mounting kit lightly into the hole opening on the top side of the housing The flap spr...

Страница 14: ...cover has two suspension openings and two additional prepared fastening possibilities3 4 Insert the batteries at the back of the device and remove the insulation strips Make sure they are inserted correctly The poles are marked on the battery and in the battery compartment 3 Only the mounting points provided should be used for mounting ...

Страница 15: ...he large rubber seal use the mounting cone in AC1321 pass the cable through it and fit the rubber plug in the cover 6 Only for LAN devices Plug the cable into the device4 For secure fastening and to ensure a reliable data link make sure it clicks in audibly 4 The connected cable may not exceed a length of 30 m at most Disruptions can occur in operation if a longer cable is used ...

Страница 16: ... power supply 8 The wall bracket is fastened to the wall with screws according to the drilling template The screws may only protrude so far that the device can click into place properly when put on to the fastened cover 56 5 The screws must not be tightened 6 Only use the screws provided in the package Screw specifications M3 5 head strength 2 5 mm head diameter 7 mm ...

Страница 17: ...x Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 17 of 35 9 The housing is put in the wall bracket and clicked into place 10 The device can then be integrated into the server software The procedure is described in the manual E SM RMS WEB ...

Страница 18: ...RMS LOG Wireless LAN Data Loggers E M RMS LOG V1 4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 18 of 35 11 Insert the probe and fasten with the thumb screw ...

Страница 19: ...G Wireless LAN Data Loggers E M RMS LOG V1 4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 19 of 35 3 1 Drilling Template Wall Bracket Figure 4 Drawing of the wall bracket not to scale ...

Страница 20: ...als Otherwise the device could be damaged 4 1 Battery The batteries serve to supply the device with power in the event of a failure in the external power supply The functionality of the device is restricted in battery mode The device continues to measure and records all data in the internal memory The device cannot communicate via the Ethernet interface but the wireless interface works normally Li...

Страница 21: ...onfiguration is obtained automatically RMS WEB URL Host rms rotronic com RMS WEB URL Path wService wService3 DEviceService svc Wireless Devices The device does not have a wireless configuration on leaving the factory It is assigned automatically when pairing the device in a system 5 2 Configuration of the LAN Devices with RMS CONFIG If you do not want to connect the device to the Rotronic Cloud th...

Страница 22: ...n configured with the correct server address the devices can then be integrated into the server software Details are described in the manual E SM RMS WEB 5 3 General Manipulations Remove Ethernet Cable The cable can be released with a blunt round object e g ballpoint pen and the connector pulled out see Figure 6 Figure 6 Removing the Ethernet cable ...

Страница 23: ...re is no DHCP server available in the network Integration of the Data Logger Pairing in 6 Steps 1 If you do not want to connect the device to the Rotronic Cloud the server must be configured in the device Connect the device to the local network and start the RMS configuration software Search for the device under Device Search Network Device The software finds all RMS devices in the local network E...

Страница 24: ...s LAN Data Loggers E M RMS LOG V1 4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 24 of 35 4 Press the button on the device briefly The device stops flashing 5 Configure the device The RMS D L ...

Страница 25: ...nge The search for the device is started by pressing a button If it is possible for the device to connect with the gateway this is shown by orange flashing Pairing is then completed as described in the RMS WEB software Integration of the Wireless Data Logger Pairing in 5 Steps 1 Log into the RMS software Cloud Select Extras Setup Devices New Wireless Device 2 Select the gateway you want your wirel...

Страница 26: ...anual 2018 ROTRONIC AG Bassersdorf Switzerland Page 26 of 35 4 Configure the device 5 Finish configuration You can find details in the instruction manual for the RMS server software E SM RMS WEB 5 5 Function Overview Overview of the main software functions of the device ...

Страница 27: ...ftware Host Address of the server with the RMS software Server path Server path where the server software is installed Measurement by the connected probe The device recognizes the connected probe automatically and requests the latest measured data The measured data are sent to the RMS server software at the set interval directly after the measurement Save measured data The measured values of every...

Страница 28: ... of RMS devices typically last 3 years The device shows automatically when the battery needs to be replaced LED flashes red System message in the RMS server software The following steps are necessary to replace the battery Take the device out of the wall bracket Remove the old battery and insert a new one The time setting of the data logger is synchronized automatically after the battery replaceme...

Страница 29: ... RMS LOG V1 4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 29 of 35 7 Firmware Update The firmware can be updated with the RMS server software Firmware updates are available for downloading on the Rotronic website ...

Страница 30: ...L Wireless 868 MHz RMS LOG 868 Wireless 915 MHz RMS LOG 915 Protocols HTTP MODBUS Wireless range 20 50 m indoors 868 MHz 15 25 m indoors 915 MHz Transmitting power 14dBm 25mW 2dBm 1 6mW Software compatibility V1 1 V1 2 1 General Device type RMS Display IP protection class IP658 Range of application 20 30 C 0 100 RH Storage and transport conditions 20 30 C 0 100 RH Interfaces Ethernet RMS D L Proto...

Страница 31: ...al Wireless logger 1 s typical LAN display 20s Measurement interval 10 s to 15 min Refresh rate RMS D L 10 s Housing Specifications Housing material ABS Dimensions 105 x 113 x 38 mm Weight 200 g Conformity EMC directives RMS LOG L EMC Directive 2014 30 EU LVD Directive 2014 35 EU EN 61326 1 2013 Industrielle Umgebung IEC 61326 1 2012 ed2 0 Industrial Environment EN 55011 2016 class B IEC CISPR 11 ...

Страница 32: ...rmance criterion www rotronic com EMC directives RMS D L EMC Directive 2014 30 EU LVD Directive 2014 35 EU EN 61326 1 2012 Industrielle Umgebung IEC 61326 1 2013 Industrial Environment EN 55011 2016 class B IEC CISPR 11 2015 class B EN 55032 2015 IEC CISPR 32 2015 EN 61010 1 2010 IEC 61010 1 2010 EN 50581 2012 Performance criterion www rotronic com FCC RMS LOG 915 FCC 47 CFR part 15 subpart B Clau...

Страница 33: ...bles adapters calibration material etc are to be found in the manual E M HC2 Accessories 9 1 RMS Accessories Order Code Description HCD S Standard probe black RH C PCD S Standard probe anthracite Differential pressure CCD S Standard probe anthracite CO2 RMS GW 868 Gateway LAN to 868 MHz AC1321 Mounting kit with Allen key and mounting cone RMS NPK Network planning kit wireless dongle RMS mini logge...

Страница 34: ... 10 Additional Documents Document Name Contents E IM RMS WEB Instruction Manual System Installation E SM RMS WEB Instruction Manual System Startup E OM RMS WEB Instruction Manual System Operation E M RMS GW 868 Instruction Manual Gateway E M RMS MLOG Instruction Manual Mini Logger E M RMS LOG R Instruction Manual Din rail logger ...

Страница 35: ...Display New 1 5 Power suppler Updated for the RMS Display 1 6 Measured parameters Updated table 1 9 Indicator and button Updated table 5 4 1 Configuration RMS D L 8 Technical specification New RMS Display details 10 Additional documentation New Manual V1 3 March 2018 1 5 1 Battery specification Updated table 1 1 Modbus Updated Read input register 04 8 Technical specification Updated table 9 1 RMS ...

Отзывы: