38
Setup
Resetting the calibration to the factory defaults
Uncheck the checkbox
4
(Apply calibration) and transfer the change to the logger
by clicking
Write user settings
.
CAUTION
If, following calibration, changes are made via
„Changing the basic unit“ on page 36, the per-
formed calibration will be lost .
Changing the display parameter
Changes that are made with
Change display parameter
have no effect on the
settings made with
Change basic parameter
. This allows you to always perform
calibration using the units selected with
Change basic parameter
.
Select the desired channel (e .g . analogue input A2) and then click
Change dis-
play parameter
. Enter your adjustments
5
(e .g . psi instead of Mbar) .
To enter the conversion factor (e .g . Mbar to psi) the following options are available:
• Enter the
amplification
and the
zero point
• Using the calculation assistant
6
enter the values for points P1 and P2
7
. The
program will then calculate the amplification and the zero point .
Activate “Apply user information” by checking the checkbox
4
and then click
OK
. Transfer the changes to the logger by clicking
Write user settings
.
Resetting the properties to the factory settings:
Uncheck checkbox
4
(
Apply user information
) and transfer the changes to the
logger by clicking
Write user settings
2
.
3
4
6
6
Max . 4 characters
7
5
Содержание LOG-HC2
Страница 5: ...User manual...
Страница 9: ...Instructions Operating Instructions LOG HC2 RO1 LOG HC2 P1...
Страница 16: ...Instructions 16 Specifications See data sheets Defining the axes Y Z Z Y X X...
Страница 20: ...Setup Online Viewer Reader PC software 20 Setup Reader Viewer Online...
Страница 50: ...50 CSV Concat Calc Cutter Utility programs...
Страница 58: ...58 FreeMat Additional software for PCs...
Страница 60: ...11 2011 LOG HC2 Firmware V1 12 LOG HC2 PC Software V2 00 60...