- 23 –
______________________________________________________________________
in_hypalm_aw1_v3.doc
SYS STATUS
Definition
This function displays the software version of the instrument, any user defined description for the
instrument (requires the optional HW3 software) and the serial number of the instrument.
Procedure
When entering the function, the software version number is displayed first. Use the UP or the
DOWN key to display the other data.
Errors and Status Messages
The following is a list of coded messages (101, etc.) that the HygroPalm may show on the bottom
line of the LC display.
Errors:
101
checksum error
the checksum test did not pass during RS-communication.
102
bad command
an unknown command was received
103
disallowed
command
a command was received that is reserved for production and
service
104
unknown probe
input
reference was made to a non-existing analog or digital probe
input
105
argument error
error in the number of the arguments in the command or in
the value of one of the arguments
106
HygroClip
communication error
the HygroClip probe does not answer or is not connected
107
calibration error
the difference between the probe reading and the calibration
point is larger than the maximum allowed by the INI
command.
108
calibration error
(overflow)
internal probe error (or the difference between the probe
reading and the calibration value is too large)
110
unknown reference
probe
the reference probe is not connected or the reference probe
input does not exist
111
Temperature error
During humidity calibration, temperature should be within the
limits of 0 and 80
q
C (32 and 176
q
F)
Warnings:
120
no adjustment
calibration
121
No HygroClip probe
is connected
Status:
130
the probe was
adjusted
calibration