Humi Unit:
use the UP or the DOWN key to change the symbol used for relative humidity
(depends on country and language).
7.8 SYS Status
Definition
This function displays the software version of the instrument, the RS232 configuration * (baud rate,
parity, bits and stop bit), the network address *, any user defined description for the instrument *
and the serial number of the instrument.
This information is for display only. Some parameters such as the network address can be
changed with the HW4 software, after connecting the instrument to a PC.
* except HygroFlex 1
Procedure
When entering the function, the software version number is displayed first. Use the UP or the
DOWN key to display the other data.
7.9 OUTPUT
Definition
This function displays the numerical values sent to the analog outputs. The display shows the
values at the time the menu key was pressed and these values are not updated to the display as
long as the function is active. This allows the values to be correctly compared.
Note: the analog output signals keep being updated while this function is in use.
Procedure
When entering the function, analog output 1 is displayed first. Use the UP or the DOWN key to
display the other outputs.
8
Error and status messages
The following is a list of coded messages (101, etc.) that the HygroFlex may show on the bottom
line of the optional LC display or via PC interface.
Errors:
101
checksum error
The checksum test did not pass during RS-communication.
102
bad command
An unknown command was received
103
disallowed command
A command was received that is reserved for production
and service
104
unknown probe input
Reference was made to a non-existing analog or digital
probe input
105
argument error
Error in the number of the arguments in the command or in
the value of one of the arguments