Figure 2. An example of Automatic Zero Point Calibration 

The Automatic Zero Point Calibration should be executed after installation and at least once a 
year.  

Trouble shooting 
The yellow LED does not start to flash. The sensor does not enter calibration mode within three 
minutes. 

1.

The absorber of the zero gas bag is not fresh. Remove the old absorber and refill the bottle with
fresh. 

The absorber is corrosive! Be careful and wear eye protection goggles

.

2.

The flow of the zero gas is too small or air is leaking into the sensor. The gas flow must be about
0,5 l/min.

3.

Push the push button on the PCB for ten seconds (until the red LED is lit) to force manual zero
point calibration. Please check the gas before pushing the push button since the calibration is
executed even if the gas is not good enough.

Contact the dealer if the sensor still does not enter the calibration mode. 

0

100

200

300

400

500

600

700

800

0

50

100

150

200

250

300

350

p

p

m

 C

O

2

Time seconds

Normal 

measurement

Zero gas 

Yellow LED 
flashes

Automatic Zero 
Point Calibration 
is reached

Phase 1 Check 
calibration conditions

Phase 2 Acknowledgement period

Calibration is executed The yellow 

LED flashes four times The red LED 

is lit The relay is closed

上海萨海测量技术有限公司  
电话:021-6236 2960
传真:021-5235 2321
邮箱:[email protected]
网址:www.seasy-ist.com

Содержание CF8-D

Страница 1: ...A 148 Rev 1 Page 1 5 User manual CF8 D W IN General The sensor CF8 D W IN is used to measure the carbon dioxide concentration inside incubators All functions can be modified from a PC with the communication cable ...

Страница 2: ... concentration OUT2 measuring signal of carbon dioxide concentration Terminal Standard configuration Standard settings Standard function OUT1 4 20mA 0 3 CO2 Measuring signal OUT2 0 5VDC 0 2 CO2 Measuring signal Table I Default configuration of CF8 D W IN OUT1 Current Voltage Connection in position Current provides 4 20mA output range for OUT1 See note 1 Connection in position Voltage provides 2 10...

Страница 3: ... sensor by power OFF ON Contact local distributor 1 2 Reserved 2 4 Algorithm Error Indicate wrong EEPROM configuration Try to restart sensor by power OFF ON Check detailed settings and configuration with UIP software version 4 0 and higher Contact local distributor 3 8 Output Error Detected errors during output signals calculation and generation Check connections and loads of outputs Check detaile...

Страница 4: ...ithin eight minutes the sensor returns to normal measuring operation without calibration 4 When the calibration is successfully executed the yellow LED flashes four times pauses and flashes another four times for an eight minutes acknowledgement period phase 2 During the acknowledgement period the relay is closed and the red LED is lit 5 Remove the zero gas during the acknowledgement period The se...

Страница 5: ...ow must be about 0 5 l min 3 Push the push button on the PCB for ten seconds until the red LED is lit to force manual zero point calibration Please check the gas before pushing the push button since the calibration is executed even if the gas is not good enough Contact the dealer if the sensor still does not enter the calibration mode 0 100 200 300 400 500 600 700 800 0 50 100 150 200 250 300 350 ...

Отзывы: