Automatic Zero Point Calibration 

The 

CF8-D/W-IN

 sensors have a function for easy automatic zero point calibration. The calibration 

should be done after installation and at least once every year. The calibration can be done on site. It is 
only necessary to connect the zero gas without any further action required from the operator, but to 
check the verification signal offered from the system. 

1.

Unscrew one of the filters from the sensor housing and replace it with a hose nipple.

2.

Connect the zero gas to the nipple. The sensor automatically enters phase 1 of the calibration
procedure when the CO

2

 concentration drops sharply. See figure 2. Do not remove the zero gas

during calibration.

3.

When the calibration starts the yellow LED starts to flash. The sensor continues the procedure
until the readings of the sensor are stable enough and are below the limit of Automatic Zero Point
Calibration. If the conditions are not good enough for calibration within eight minutes the sensor
returns to normal measuring operation without calibration.

4.

When the calibration is successfully executed the yellow LED flashes four times, pauses and
flashes another four times for an eight minutes acknowledgement period, phase 2. During the
acknowledgement period the relay is closed and the red LED is lit.

5.

Remove the zero gas during the acknowledgement period. The sensor returns to normal
measuring mode as soon as the calibration is ready. The yellow LED continues flashing, the red
LED is lit and the relay is closed independent of gas removal.

The zero gas can be generated by passing air through the chemical absorber in the service bag 
F0005 or by using nitrogen in a gas bottle. Not more than three minutes shall pass before the yellow 
LED starts to flash. If the yellow LED does not start to flash the zero gas is not good enough or the 
sensor is not stable enough. Wait ten minutes before another attempt to calibrate the sensor.  

If the sensor does not enter phase 1 even if the conditions are good push the push button on the PCB 
until the red LED is lit to force a manual zero point calibration. Push the push button for at least ten 
seconds. 

Содержание CF8-D

Страница 1: ...A 148 Rev 1 Page 1 5 User manual CF8 D W IN General The sensor CF8 D W IN is used to measure the carbon dioxide concentration inside incubators All functions can be modified from a PC with the communication cable ...

Страница 2: ... concentration OUT2 measuring signal of carbon dioxide concentration Terminal Standard configuration Standard settings Standard function OUT1 4 20mA 0 3 CO2 Measuring signal OUT2 0 5VDC 0 2 CO2 Measuring signal Table I Default configuration of CF8 D W IN OUT1 Current Voltage Connection in position Current provides 4 20mA output range for OUT1 See note 1 Connection in position Voltage provides 2 10...

Страница 3: ... sensor by power OFF ON Contact local distributor 1 2 Reserved 2 4 Algorithm Error Indicate wrong EEPROM configuration Try to restart sensor by power OFF ON Check detailed settings and configuration with UIP software version 4 0 and higher Contact local distributor 3 8 Output Error Detected errors during output signals calculation and generation Check connections and loads of outputs Check detaile...

Страница 4: ...ithin eight minutes the sensor returns to normal measuring operation without calibration 4 When the calibration is successfully executed the yellow LED flashes four times pauses and flashes another four times for an eight minutes acknowledgement period phase 2 During the acknowledgement period the relay is closed and the red LED is lit 5 Remove the zero gas during the acknowledgement period The se...

Страница 5: ...ow must be about 0 5 l min 3 Push the push button on the PCB for ten seconds until the red LED is lit to force manual zero point calibration Please check the gas before pushing the push button since the calibration is executed even if the gas is not good enough Contact the dealer if the sensor still does not enter the calibration mode 0 100 200 300 400 500 600 700 800 0 50 100 150 200 250 300 350 ...

Отзывы: