A2: AwTherm Specifications
General
Device type
Temperature-stabilized water-activity measurement
Operating conditions
1..40 °C (34..107 °F)
Measurement range
0.005…1.000 aw
Accuracy
±0.005 aw
±0.1 °C (±0.18 °F)
Power supply
110..230 V / 50..60 Hz
Display
8-line LCD with touch operation
Sample sizes
Variable (14 mm / 40 mm)
Technical Information
Current Consumption
≤
2 A
Temperature-control range
0..60 °C (32..140 °F)
Temperature stability
±0.01 °C/min (±0.018 °F/min)
Chamber-temperature gradient
<0.1 °C (<0.18 °F)
Firmware update
Via USB port
Functions
HW4-compatible
Yes (v 3.6.0 and higher)
aw-Quick function
Yes
Interface
Micro USB
Calibration / Adjustment
Aw & Temperature:
HG2-S (probe removed with AwT-CAL)
Aw:
ROTRONIC humidity standards (via HW4)
Trend indicator
Yes
Probes
Sensor
HYGROMER
®
IN-1
Maintenance / Calibration
Annual calibration (recommended)
Long-Term Stability
<0.01 aw/year
Temperature sensor
Pt100, DIN 1/3 Class B
Approval / Conformity
Standards
ISO 21807
CE / EMC
EMC 2004/108/EC
IEC/ EN 61010-1:2010
IP protection
IP21
Housing / Mechanical parts
Housing material
PC / ABS
Housing dimensions
400 x 180 x 180 mm
Sample container holder dimensions
AwT-PS14: Ø46 x 14 mm
AwT-PS40: Ø46 x 40 mm
Weight
4200 g