background image

2.8 Frequently asked questions  

Question: 

I am using at present a ACRL010 on a HygroPalm 3, input 1. I plug into the M-series. Input 2 is 
connected to a MOK-02-B5 with my reference HygroClip. I read the unit and compare the value with the 
value of the HygroClip on input 2. I go then to adjust the M22 series. I go to adjust REF. Palm displays 
“Palm = Reference” and I press OK, and I adjust my M-series. Is this OK or do I risk errors on the ACRL 
010 ?  The input 1 of the HygroPalm is configured for a digital probe.

 

 
Answer:  

The HygroClip probes connected on any of the M-series transmitters can be adjusted that way. It 
must be pointed out, that only the HygroClip probes, but not the output signals can be adjusted 
that way. Of course, the output signal will be changed also according to the adjustment deviation 
of the probe. 

 
Question: 

When I do a multi point adjust, I do e.g. 35 %  as first point and I come out of the menu and change my 
solution to e.g. 80 %. I then await stabilization and go back into the adjust menu and do my 80 %. How 
does the clip know that its not a 1 point adjust at 80 % instead of a second point at 80 %? The fact that I 
began at 35 % does it memorize  or does this set something off ? 

 
Answer: 

The determination is done in the adjustment window of HW3. There you select either a single or 
multi-point adjustment. 

 

If a multipoint adjustment is done, the software will check for the entered values.If a value 
between 25 and 55 %rh is entered, the algorithm will consider this as the offset point. A value >55 
% is considered as slope value, while a value >1>25 % is used for linearity adjustment in the low 
range. 

 
Question: 

When I do a 1 point adjustment, is the adjusted probe only correct for the point done and false for the rest 
of the curve.? When I do one point do I not just displace the whole slope up or down? 

 
Answer: 

A one point adjustment always just influences the offset, but not the slope and neither the 
linearity. This means that on the particular point, the instrument will be accurate, but the farther 
away from that point, the bigger the deviation. Within HW3 and also with HygroPalm handhelds, 
you can always select the mode of adjustment: 1 point or multiple point. When you select one 
point, you definitely move the offset and only the offset. 
 
The offset is the parallel movement of the whole curve in vertical direction as shown in the left 
graph, while the slope is the steepness of the curve, as the graph to the right shows. 
 
 
 
 
 
 
 
 
 
 
       Offset 

 

 

 

Slope 

 
 
A fourth point may be adjusted in order to get perfect linearity.  
Details see in subsequent pages. 

\\Kalibrierung\Calibration_procedures_2.doc 

9

Содержание ACML232

Страница 1: ...umidity Standard 3 2 Cable Overview 5 2 1 HygroPalm connected to PC 5 2 2 HygroPalm connected to HygroFlex 5 2 3 HygroPalm connected to M Roline L Series 6 2 4 M23 M33 connected directly to PC 6 2 5 M...

Страница 2: ...or water dripping water attempts to measure non hygroscopic substances The reproducibility and long term stability are not influenced even if the sensor was exposed to high humidity or saturation wit...

Страница 3: ...NTER The display shows SURE Press enter to confirm and adjust Now the reference adjustment is finished Note If you use a calibration device EDM15 15 you can insert both probes at the same time to ensu...

Страница 4: ...andards includes 5 discs in the receptacle of the calibration device The purpose of this disc is to prevent accidental spilling of the solution inside the calibration device or on the humidity sensor...

Страница 5: ...2 is always the reference probe 2 2 HygroPalm connected to HygroFlex Service Cable Palm 2 Palm 3 Possible Functions yes yes Display the readings from any probe connected to the HygroFlex yes yes Displ...

Страница 6: ...int adjustment of any digital probe respectively transmitter using a RHS humidity standard yes yes yes yes Multi point adjustment of any digital probe respectively transmitter using a RHS humidity sta...

Страница 7: ...ted to PC via MOK WIN and ACRLXB5 Device M1xW M1xD L1xD W Possible Functions Accessories required Yes Yes Display the measurement values of the M1 L1x transmitter Yes Yes 1 point adjustment using a RH...

Страница 8: ...point or 2 point adjustment of temperature Yes Yes Display the serial number last calibration date of temperature and humidity calibration PC running HW3 software AC1625 MOK 02 WIN eventually AC adapt...

Страница 9: ...r The determination is done in the adjustment window of HW3 There you select either a single or multi point adjustment If a multipoint adjustment is done the software will check for the entered values...

Страница 10: ...a linearisation The Hygromer sensor is by its nature linear between 30 and 100 rh Below 30 rh there is a kink in the curve which needs linearisation Therefore the adjustment point are set as follows...

Страница 11: ...e is not going trough zero and maximum the slope and linearity must be adjusted next The slope will be set at 80 This results in linearity between 30 and 100 0 10 35 80 Output Ideal curve Before adjus...

Страница 12: ...at 0 0 5 rh When all these adjustments are done the characteristic line is linear and the probe is perfectly adjusted 0 10 35 80 Fix due to 35 adjustment Output Fix due to 80 adjustment Relative humi...

Страница 13: ...ope and linearity are not adjusted the ranges that are far away from the adjusted point are not very accurate We therefore recommend to perform multiple point adjustments in order to achieve maximum a...

Отзывы: