140024 1-4_ D14.746.19.03_2019-07-04 QuattroPlus Eng
Page
7
of
12
•
A basic cleaning with our special cleaner (order no. 1400677) or isopropyl alcohol will increase the suction
effect and is recommended in larger time intervals (dependent on the frequency and conditions of use, e.g. after
use of bath additives containing oil).
Use these cleaning agents only on the rubber pads but never on the plastic parts.
•
Disinfect the product by wiping all surfaces with a cloth with disinfectant.
•
In a medical environment, also refer to the "Recommendations of the Commission for Hospital Hygiene and
Infection Prevention" on the website of the Robert Koch Institute www.rki.de.
•
The grab handle is resistant to the active agents in the hand and surface disinfectants certified by the DGHM
(Disinfection Commission of the German Society for Hygiene and Microbiology) with the exception of phenol
ether, pyridine derivative, and thiocyanate.
•
Only alcoholic disinfection without re-fattening component may be used on the rubber discs.
5.2 Maintenance
(service):
•
If the handle has been attached to the wall for a prolonged period of time, the rubber discs will overstretch. We
therefore recommend that you remove the handle from the wall approx. every 14 days and leave it overnight on
a flat surface with the rubber disks relaxed, then reattach it the next morning.
•
Examine the rubber pads for possible functional deterioration because of reduced elasticity due to age, wear,
exposure to sunlight, uneven storage, scoring, etc.
If the rubber pads are damaged and therefore the adhesive strength is limited, the complete suction heads
(spare part no. 14002S) are to be replaced or the rubber pads (spare part no. 1402721) must be exchanged by
trained specialist staff.
•
Keep the vacuum safety indicator always free from dirt!!
•
Check the handle regularly for damage and do not use the handle under any circumstances if it is
damaged.
6. Recognizing and remedying malfunctions
Malfunctions
Possible causes of
malfunctions/test
Corrective measures
With new handle: safety
indicator pops up after a
short time
Lime or other dirt on the surface.
Clean the surface thoroughly.
The surface is not non-porous even if it
looks smooth, or it is damaged (e.g. hairline
crack in the tile).
Possible test procedure: Attach the handle
to a glass pane or mirror and pay attention
to the safety indicator.
If now the safety indicator does not move
any more, the fault is in the substrate.
If possible, attach the handle at another location
where the underground is non-porous.
If this is not possible, self-adhesive stainless-steel
plates (order no. 1461418) to which the handles can
be attached can be glued on. See accessories.
With older handle: safety
indicator pops up after a
short time
Rubber discs scratched or damaged.
If there is damage, the rubber discs (order no.
1402721) can be replaced by a specialist dealer or
the complete suction head (order no. 14002S) can
be replaced.
Indicator rocker is still
slightly raised after
mounting
1. Toggle lever not firmly pressed down
before turning over!
Attach the handle again while beforehand pressing
the rocker arm firmly down. (Page 8 fig. 2a arrow 2)
2. The rubber pad is overstretched.
This happens when the handle has
been attached to the wall for a long
time.
Leave the handle overnight on a flat surface with the
rocker released (page 9 fig. 12). Attach again the
next day while beforehand pressing the rocker arm
firmly down. If the mounting rocker still protrudes
after pressing it in, the rubber discs (order no.
1402721) must be replaced by a specialist dealer or
the complete suction head (order no. 14002S) must
be replaced.
7. Disposal:
The product packaging should be recycled. The metal parts can be recycled as scrap metal. Disposal must be
carried out according to the respective national statutory regulations. Please ask your local or communal authority
about local disposal facilities.
These ROTH mobile QuattroPlus grab handles may not be exported to the U.S. (cf. U.S. version).
Subject to technical modification.