EMC guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The MS400i is intended for use in the electromagnetic environment specified below.
The customer or the user of the MS400i should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The MS400i uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The MS400i is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic pur-
poses.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Compliance
Guidance and manufacturer’s declaration-electromagnetic immunity
The MS400i is intended for use in the electromagnetic environment specified below.
The customer or the user of the MS400i should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guid-
ance
Electrostatic
discharge(ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast tran-
sient/burst IEC 61000-
4-4
± 2kV for power supply
lines
± 1kV for input / output
lines
± 2kV for power supply
lines
Not applicable
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Surge IEC 61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
± 1kV differential mode
Not applicable
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip in UT)
for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in UT)
for 5 s
<5% UT(>95% dip in UT)
for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in UT)
for 5 s
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the MS400i requires
continued operation during power mains
interruptions, it is recommended that the
MS400i be powered from an uninterrupt-
ible power supply or a battery.
Power frequency
(50/60 Hz) mag-netic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristics of a typical lo-
cation in a typical commercial or hospital
environment
NOTE: UT is the a.c. mains voltage prior to application of the test level.