background image

User Manual 

 

RPC-2.92  

Calibration Kit

  

LRL-Version

 

02CK020-150 

 

 

Rosenberger Hochfrequenztechnik GmbH & Co. KG  
P.O.Box 1260       D-84526 Tittmoning       Germany 
www.rosenberger.com

 

Tel.    : +49 8684 18-0  
 
Email : [email protected]

 

Page 

 

1 / 20

 

 

D

ie

s

e

s

 D

o

k

u

m

e

n

is

u

rh

e

b

e

rr

e

c

h

tl

ic

h

 g

e

s

c

h

ü

tz

 T

h

is

 d

o

c

u

m

e

n

is

 p

ro

te

c

te

d

 b

y

 c

o

p

y

ri

g

h

 R

o

s

e

n

b

e

rg

e

H

o

c

h

fr

e

q

u

e

n

z

te

c

h

n

ik

 G

m

b

H

 &

 C

o

K

G

 

 

 

 

 

 

 

02CK020-150 (LRL Version) 

Содержание 02CK020-150

Страница 1: ...echnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 20 Dieses Dokument ist urheberrechtlich gesch tzt This document is prot...

Страница 2: ...umentation 1 4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installation of...

Страница 3: ...d care VNAs usually offer several techniques for correcting systematic measurement errors The standards in this calibration kit allow you to perform two port LRL and TRL calibrations with best perform...

Страница 4: ...wrench 03W008 000 5 mm 6 mm 7 mm 8 mm 9 mm 1 Torque wrench 03W021 000 8 mm wrench size 0 9 Nm torque 1 1 4 Kit Documentation USB Stick Standard Definitions as data files for Vector Network Anaylzer F...

Страница 5: ...the allowable temperature drift to 1 K during measurement calibration and during measurements when the network analyser error correction is turned on For further information please refer to the VNA s...

Страница 6: ...del is used with a short inductance L0 to L3 of 0 H Short calibration standards OSL OSM and OSLT TOSM calibrations which can only be carried out with additional calibration standards need well modelle...

Страница 7: ...complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously loading...

Страница 8: ...vector network analyzer Main setting parameters It is good practice to start your measurement by selecting the main setting parameters Sweep type linear logarithmic Sweep range start frequency stop fr...

Страница 9: ...ou thru the settings and actions For details please consult the manual of your specific VNA 4 2 Calibration Check It is advisable to check the calibration after error correction has been performed par...

Страница 10: ...be concatenated to several frequency segments to allow a wider frequency range Please consult your VNA operating manual TRL Calibration Thru Line Min frequency Max frequency T L1 15 0 mm 1 1 GHz 8 9 G...

Страница 11: ...se of re calibration or repair contact your local Rosenberger sales office 5 4 Extent of Rosenberger Re Calibration Services Every calibration kit sent for re calibration or repair the following servi...

Страница 12: ...ector repeatability and its effect on the overall uncertainty of a measurement procedure should encourage careful inspection interface gauging and handling of coaxial connectors Prior to use a visual...

Страница 13: ...ost importance that connectors do not damage the test equipment interfaces to which they are connected Poor performance of many coaxial devices and cable assemblies can often be traced to poor constru...

Страница 14: ...der are vice versa Checking devices for protrusion or excessive recession only a single measurement is sufficient More precise measurements or checking devices very close to critical limits should be...

Страница 15: ...ent ist urheberrechtlich gesch tzt This document is protected by copyright Rosenberger Hochfrequenztechnik GmbH Co KG 6 2 4 Gauge blocks Every connector gauge requires a gauge block that is used to ze...

Страница 16: ...ch the centre conductor extends beyond the outer conductor mating plane This may result in buckling of the socket contacts or damage to the internal structure of a device due to the axial forces gener...

Страница 17: ...to 4 1 Connector Care 4 Use a connector gage to verify that all centre conductors are within the observed pin depth values in Table 2 2 Refer to 4 2 Interface Dimensions and Gaging 5 Carefully align t...

Страница 18: ...connectors Connecting Air Lines This calibration kit comes with unbeaded Air Lines This means that the outer conductors do not hold the inner conductors resulting in the best Air Line performance but...

Страница 19: ...er the inner conductor without touching it Tighten the coupling nut with your fingers Hold the test port adapter with the appropriate wrench size of the combi wrench Use the torque wrench to fasten th...

Страница 20: ...ll the connectors straight apart without twisting rocking or bending either of the connectors The inner conductor of the Air Line stays on that side with the higher connector forces So it may keep ins...

Отзывы: