![Rosemount 7D Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/rosemount/7d/7d_instruction-manual_1487404016.webp)
I
NTRODUCTION
4
December 2000 Rosemount Analytical 748221-L
Model 7D Thermal Conductivity Analyzer
1.2 GAS SELECTOR PANEL
If so ordered, the analyzer module is provided with an appropriate gas selector panel,
Figure 1-3. The gas selector panel permits selection, flow adjustment, and flow
measurement for the various gases: sample; flowing reference gas, if used; and
downscale and upscale calibration gases. Proper choice of a gas selector panel
depends on:
1. Configuration of the thermal conductivity cell, i.e., flowing or sealed-in reference
gas.
2. Composition of the sample stream. For non-corrosive streams, the gas selector
panel is assembled with brass components. For corrosive streams, stainless steel
is used.
F
IGURE
1-3. T
YPICAL
G
AS
S
ELECTOR
P
ANEL
DESCRIPTION
PART
NUMBER
Brass and Copper construction for use with sealed reference
113357
Stainless steel construction for use with sealed reference
113920
Brass and Copper construction for use with flowing reference
117195
Stainless steel construction for use with flowing reference
118210
T
ABLE
1-1. A
VAILABLE
G
AS
S
ELECTOR
P
ANELS
DOWNSCALE
CAL GAS
UPSCALE
CAL GAS
SAMPLE
REFERENCE
SAMPLE
REF
1
Reference Gas
Flow Meter
Sample/Calibration
Gas Flow Meter
Downscale
Calibration Gas
Needle Valve
Upscale
Calibration Gas
Needle Valve
Sample Gas
Needle Valve
1
Reference Gas
Needle Valve
1
Provided only if thermal conductivity cell uses flowing reference gas.
Содержание 7D
Страница 12: ...PREFACE P6 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 30: ...INSTALLATION 18 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 34: ...INITIAL STARTUP 22 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 36: ...OPERATION 24 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 54: ...REPLACEMENT PARTS 42 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...