![Rosemount 7D Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/rosemount/7d/7d_instruction-manual_1487404015.webp)
I
NTRODUCTION
748221-L Rosemount Analytical December 2000
3
Model 7D Thermal Conductivity Analyzer
Alarm Kit
±
15V Power Supply
Linearizer Board
Isolated Current Output Board
Bridge Power Supply
Fuse
Cell Enclosure
REF -
Hinge Side
Temperature Control Assembly
Master Board
AC Power Input
Terminal Block TB6
1.1.2 E
LECTRONIC
C
IRCUITRY
The analyzer module contains solid-state circuitry that conditions the bridge-imbalance
signal as required to provide readout on the front-panel meter. In addition, a
field-selectable output for a voltage-type recorder is provided as standard. A
field-selectable output of 4 to 20 mA for a current-actuated recorder or other device is
obtainable through use of an optional plug-in circuit board. A calibration curve can be
used to convert meter or recorder readings into concentration values. Typical
calibration curves are supplied for standard ranges. Calibration curves for special
ranges are available as options.
To avoid use of a calibration curve in an application where it would otherwise be
required, the analyzer may be equipped with an optional linearizer board. If so, the
linearizer is factory set for a given range only, and is not usable on another range.
Note that a linearizer is usable only if nonlinearity at midscale does not exceed
approximately 20% of fullscale.
F
IGURE
1-2. M
ODEL
7D M
AJOR
C
OMPONENT
L
OCATIONS
Содержание 7D
Страница 12: ...PREFACE P6 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 30: ...INSTALLATION 18 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 34: ...INITIAL STARTUP 22 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 36: ...OPERATION 24 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...
Страница 54: ...REPLACEMENT PARTS 42 December 2000 Rosemount Analytical 748221 L Model 7D Thermal Conductivity Analyzer NOTES ...